Recall Z-1018-2022

Becton Dickinson & Co. · initiated March 9, 2022

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

BD Synapsys Microbiology Informatics Solution, Catalog Number 444150, Software Version 4.10

Codes / lots: All systems running BD Synapsys version 4.10; UDI: 00382904441500; Serial Numbers: SL00530 SL00515 SL00527 SL00502 SL00540 SL00564 SL00549 SL00509 SL00003 SL00004 SL00010 SL00001 SL00005 SL00002 SL000015 SL01010 SL01022

Reason for recall

When updating BD Kiestra customers to BD Synapsys version 4.10, the custom antibiotic codes set by the customer are overwritten with the BD default antibiotic codes.

FDA-determined cause: Under Investigation by firm

Action taken

A MEDICAL DEVICE CORRECTION notification letter dated 3/9/22 was distributed to customers. PLEASE TAKE THE FOLLOWING ACTIONS: 1. Ensure the contents of this notification are read and understood by those within your organization. 2. BD will be contacting you within one week to request remote access to your database to confirm your configuration and evaluate potential impact. If BD determines that your facility has an affected configuration, BD will work with your facility to resolve the issue. In the interim, you many continue use of the product until BD contacts your facility. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form so that BD may acknowledge your receipt of this notification per FDA requirements. 4. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787 Actions Taken by BD: 1. BD will review configuration and evaluate potential impact for BD Kiestra customers. 2. Corrective actions have been initiated to prevent recurrence of the identified root cause. If you require further assistance, please contact:

Details

Recall number
Z-1018-2022
Initiated
March 9, 2022
Terminated
January 19, 2024
Product code
JQP – Calculator/Data Processing Module, For Clinical Use
Quantity
15 systems
Distribution
Worldwide distribution US Nationwide distribution in the states of CA, IL, MI, NY, OH, SC, and TN. The countries of Austria, France, Germany, Netherlands, Sweden.
Recalling firm
Becton Dickinson & Co., Sparks, MD

Official FDA recall record