Recall Z-1024-2015

Beckman Coulter Inc. · initiated December 15, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Beckman Coulter, Uric Acid Assay, Catalog No. OSR6X98

Codes / lots: All Lots

Reason for recall

N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations for the treatment of acetaminophen overdose, interferes with the assays for these analytes: Cholesterol OSR6X16, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30. NAC interference may lead to falsely low results.

FDA-determined cause: Under Investigation by firm

Action taken

A customer notification letter was sent to their customers on 12/15/14 to inform them that Beckman Coulter has recently confirmed that N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations (for the treatment of acetaminophen overdose), has been determined to interfere with assays for the following analytes: Cholesterol OSR6X16, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with questions are instructed to call (800) 223-0130 in the US and Canada. Customers outside the US and Canada are instructed to contact their local Beckman Coulter Representative.

Details

Recall number
Z-1024-2015
Initiated
December 15, 2014
Posted by FDA
January 28, 2015
Terminated
February 19, 2016
Product code
KNK – Acid, Uric, Uricase (Colorimetric)
Quantity
32,159 units total (10,462 units in US)
Distribution
Nationwide in US and Puerto Rico. World Wide: Albania, Algeria, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyrus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Eritrea, Ethiopia, Finland, France, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guam, Honduras, Hong…
Recalling firm
Beckman Coulter Inc., Brea, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1023-2015 Class IIBeckman Coulter, Lipase Assay, Catalog No. OSR6X30N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations for the treatment of acetaminophen… 12/15/2014Terminated
Z-1025-2015 Class IIBeckman Coulter, Cholesterol Assay, Catalog No. OSR6X16N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations for the treatment of acetaminophen… 12/15/2014Terminated
Z-1026-2015 Class IIBeckman Coulter, Lactate Assay, Catalog No. OSR6X93N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations for the treatment of acetaminophen… 12/15/2014Terminated