Recall Z-1034-2007

Abbott Laboratories · initiated April 23, 2007

Terminated

Product

AxSYM Anti-HCV Reagent Pack; list 05C36; 100 test pack; in-vitro diagnostic; Abbott Laboratories, Abbott Park, IL 60064

Codes / lots: list 05C36; lots 49454M200, 50006M100, 51129M100, 51130M100

Reason for recall

Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.

FDA-determined cause: Other

Action taken

Abbott initiated a field correction recall by letter dated 4/23/07 to all customers receiving the affected reagents. The accounts were informed of the failure of the AxSYM to open (actuate) some reagent packs with the affected lots. The accounts were requested to inspect each reagent pack for broken or improperly sealed caps; manually open each bottle to ensure the flipper bar raises the reagent cap correctly; close each flipper bar prior to loading the reagent pack on the instrument; and discard any reagent packs that do not open properly. If the user observes a probe crash, prior to testing samples again, the user must perform a probe crash recovery following the procedure in the Operations Manual. If necessary, they need to replace the probe. If a probe crash recovery is not performed, subsequent results may be impacted. The customers were requested to retain a copy of the letter for their records. Customers requiring additional information were directed to contact Abbott Customer Service at 1-877-422-2688.

Details

Recall number
Z-1034-2007
Initiated
April 23, 2007
Posted by FDA
July 11, 2007
Terminated
May 18, 2008
Product code
MZO – Assay, Enzyme Linked Immunosorbent, Hepatitis C Virus
Quantity
4,023 packs
Distribution
Worldwide, including USA, Puerto Rico, Argentina, Australia, Brazil, Canada, Cayman Islands, Chile, Colombia, Ecuador, Germany, Hong Kong, Mexico, New Zealand, Singapore, South Korea, Taiwan, Thailand and Venezuela.
Recalling firm
Abbott Laboratories, Abbott Park, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1023-2007 AxSYM Troponin-I ADV Reagent Pack; list 02J44; 100 test pack; in-vitro diagnostic…Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can… 04/23/2007Terminated
Z-1024-2007 AxSYM Toxo IgG Reagent Pack; list 03B22; 100 test pack; an in vitro diagnostic; Abbott…Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can… 04/23/2007Terminated
Z-1025-2007 AxSYM Rubella IgG Reagent Pack; list 03B23; 100 test pack; an in vitro diagnostic…Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can… 04/23/2007Terminated
Z-1026-2007 AxSYM CA 125 Reagent Pack; list 03B41; 100 test pack; an in vitro diagnostic; Abbott…Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can… 04/23/2007Terminated
Z-1027-2007 AxSYM CA 15-3 Reagent Pack; list 03B42; 100 test pack; an in vitro diagnostic; Abbott…Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can… 04/23/2007Terminated
Z-1028-2007 AxSYM AUSAB Reagent Pack; list 03C74; 100 test pack; in-vitro diagnostic; Abbott…Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can… 04/23/2007Terminated
Z-1029-2007 AxSYM Folate Reagent Pack; list 03C81 ; 100 test pack; in-vitro diagnostic; Abbott…Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can… 04/23/2007Terminated
Z-1030-2007 AxSYM Myoglobin Reagent Pack; list 03E43; 100 test pack; in-vitro diagnostic; Abbott…Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can… 04/23/2007Terminated
Z-1031-2007 AxSYM Toxo IgM Reagent Pack; list 04B25; 100 test pack; in-vitro diagnostic; Abbott…Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can… 04/23/2007Terminated
Z-1032-2007 AxSYM Rubella IgM Reagent Pack; list 04B46; 100 test pack; in-vitro diagnostic; Abbott…Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can… 04/23/2007Terminated
Z-1035-2007 AxSYM CMV IgM Reagent Pack; list 05C78; 100 test pack; in-vitro diagnostic; Abbott…Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can… 04/23/2007Terminated
Z-1036-2007 AxSYM HAVAB-M 2.0 Reagent Pack; list 05E20; 100 test pack; in-vitro diagnostic; Abbott…Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can… 04/23/2007Terminated
Z-1037-2007 AxSYM CORE 2.0 Reagent Pack; list 08B88; 100 test pack; in-vitro diagnostic; Abbott…Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can… 04/23/2007Terminated
Z-1038-2007 AxSYM CORE-M 2.0 Reagent Pack; list 08B89; 100 test pack; in-vitro diagnostic; Abbott…Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can… 04/23/2007Terminated
Z-1039-2007 AxSYM HAVAB 2.0 Reagent Pack; list 08D21; 100 test pack; in-vitro diagnostic; Abbott…Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can… 04/23/2007Terminated
Z-1040-2007 AxSYM HBsAg Reagent Pack; list 09B01; 100 test pack; in-vitro diagnostic; Abbott…Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can… 04/23/2007Terminated