Recall Z-1036-2022

American Contract Systems, Inc. · initiated March 22, 2022

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

1) ACS Spine Pack 2) ACS Lumbar/Back Set Up Tray

Codes / lots: 1) ACS Spine Pack - Model LOSP35D, Lot 966221; UDI 00191072154613 2) ACS Lumbar/Back Set Up Tray - Model SALB85AG, Lot 973221; UDI: 00191072127280

Reason for recall

Product was sterilized with a higher than specification EO concentration.

FDA-determined cause: Process control

Action taken

A customer letter dated 04/05/2022 was issued to impacted consignees. The letter instructs the following: 1) Review your inventory and segregate affected product. This product should not be shipped out to customers. 2) Return affected product to 2610 NE Industrial Drive, Suite 220, Kansas City, MO 64120. 3) If product was already further distributed, ensure that all downstream customers are properly notified of the recall and maintain records of effectiveness. A Field Action Response Form should be completed and returned to suzanne.thomas@hyh.com, even if you do not have product inventory.

Details

Recall number
Z-1036-2022
Initiated
March 22, 2022
Product code
OJG – Neurological Tray
Quantity
72 packs
Distribution
US Nationwide distribution in the states of IL, MO, NE, and TX.
Recalling firm
American Contract Systems, Inc., Kansas City, MO

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1037-2022 Class II1) ACS Arthroscopy Kit 2) ACS Knee Kit 3) ACS Hand PackProduct was sterilized with a higher than specification EO concentration. 03/22/2022Open, Classified
Z-1038-2022 Class IIACS IR - Angio PackProduct was sterilized with a higher than specification EO concentration. 03/22/2022Open, Classified
Z-1039-2022 Class IIACS Lap Chole PackProduct was sterilized with a higher than specification EO concentration. 03/22/2022Open, Classified
Z-1040-2022 Class IIACS Cath Lab PackProduct was sterilized with a higher than specification EO concentration. 03/22/2022Open, Classified