Recall Z-1038-2012

European Custom Manufacturing B.V. · initiated April 18, 2011

Terminated

Product

Temporary Pacing Wire, BIPOLAR, Models BM, V, VE and VF, Rx Only, STERILE/EO, Single use only -- Mce (MEDICAL CONCEPTS EUROPE) --- Manufacturer ECM B.V. Oost-Om 54, P.O. Box 53, NL 5420 AB GEMERT --- Shipping or Unit Package: 20 individual units packaged per box with 6 boxes per case. --- Device Classification Name: Electrode, pacemaker, temporary. ---The device is supplied sterile and intended for single use. Used to pace the heart during or after open heart surgery. The electrodes are placed on the heart; the other end of the pacing wire is inserted through the skin of the patient with a thoracic needle. The thoracic needle can snap-off at the end where the pacing wire is attached, after snap-off the remaining pins are connected to an external pacemaker. For temporary atrial and ventrical pacing and sensing for contemplated implant duration of 7 days or less.

Codes / lots: Model BM Lot numbers 1101044-10, 1012035-10, 1012036-10, 1011128-10, 1101023-10, 1101134-10, 1102035-10, 1102061-10, 1009089-10; -- Model V lot number: 1103066-10; -- Model VE lot numbers: 1101055-10, 1101008-10; -- Model VF lot numbers: 1011092-10, 1011148-20, 1012038-10, 1101021-10, 1101136-10.

Reason for recall

The crimping operation during product production was not performed properly. As such, the pacing wire may separate from the connection pin, which causes non-functionality.

FDA-determined cause: Process design

Action taken

ECM sent an "ECMbv RECALL OF MCE TEMPORARY PACING WIRES" letter dated May 2, 2011 to all affected customers. Customers were also informed of a potential problem via e-mail and/or phone between April 18 and 21, 2011. The letter describes the product, problem, and actions to be taken by the customers. Customers were instructed to place the affected product on hold and to return the product to the firm. For questions on this matter, contact ECM or MCE Customer Service at 1-877-886-1985.

Details

Recall number
Z-1038-2012
Initiated
April 18, 2011
Posted by FDA
February 15, 2012
Terminated
February 15, 2012
Product code
LDF – Electrode, Pacemaker, Temporary
Quantity
2,120 individual units (106 boxes)
Distribution
Nationwide Distribution-including the states of AL, AR, CA, FL, GA, KY, NC, NV, NY, TX, VA, and WI.
Recalling firm
European Custom Manufacturing B.V., Gemert

Official FDA recall record