Recall Z-1041-2026

Mobius Imaging, LLC · initiated December 4, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Detachable 10in Pendant Upgrade component for the AIRO Computed Tomography (CT) X-ray System.

Codes / lots: Part No. MI-70-0128; UDI: (01) 00869346000200 (11)230920 (21) AIRO-XXXX; Serial No. 2232206183, 2232206193, 2232206203, 2232206213, 2232206223, 2232206233, 2235522143, 2235522153, 2235522163, 2236339153, 2236339163, 2236339173, 2236339183, 2303429483, 2303429503, 2313925663, 2313925673, 2316030773, 2316030783, 2316703503, 2316703523, 2317415473, 2320006923, 2320006933, 2323513413, 70-0128-13904392.

Reason for recall

AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.

FDA-determined cause: Labeling Change Control

Action taken

An URGENT MEDICAL DEVICE CORRECTION notice dated 12/4/25 instructs consignees to ensure the following mitigations are in place while the correction is deployed: the hand position warnings as indicated in the IFU, train operators, the AIRO is accessible through a physical key switch and password protected log in screen, and the Backing Up System Warning label is affixed on to the gantry of affected devices. Stryker will correct devices during an upcoming service visit. Consignees are to inform users of devices of this recall by sharing the recall notice, immediately check stock areas and operating rooms to identify affected devices, and inform all users of the potential harms and their risk mitigations. Consignees with any questions are to contact their local sales office or Stryker Representative.

Details

Recall number
Z-1041-2026
Initiated
December 4, 2025
Posted by FDA
January 9, 2026
Product code
JAK – System, X-Ray, Tomography, Computed
Quantity
26 units
Distribution
US Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DC, FL, IL, KY, MA, MD, MI, MN, MO, NC, NY, OH, PA, RI, SC, TN, TX, UT, VA, WI.
Recalling firm
Mobius Imaging, LLC, Shirley, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1040-2026 Class IIDetachable 10in Pend FRU-1 Pend component for the AIRO Computed Tomography (CT) X-ray…AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label. 12/04/2025Open, Classified