Recall Z-1045-2016

Maquet Cardiovascular Us Sales, Llc · initiated January 7, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Maquet Getinge Group 1180.19XO Extension Device. Manufacturer MAQUET GmbH, Used to support the upper and lower extremities immediately before, during and after fracture surgery. The Extension Device (1180.19F0) is a table top module for use with the Maquet MAGNUS operating table system and it is attached to the base mounting point of the MAGNUS operating table.

Codes / lots: Catalog Number 1180.19F0; Serial numbers 00003 and 00004.

Reason for recall

MAQUET Medical Systems USA is initiating a field action for correction of the MAQUET Extension Device (Part No 1180.19F0). The Extension Device is manufactured by Maquet GmbH, in Germany, and is a table top module for use with the Maquet MAGNUS operating table system. Maquet GmbH was made aware of a potential issue that can result in the Extension Device becoming unlocked from the table top, causing it to fall down.

FDA-determined cause: Under Investigation by firm

Action taken

Maquet sent "URGENT PRODUCT RECALL MEDICAL DEVICE CORRECTION" letters/MEDICAL DEVICE CORRECTION RESPONSE FORM dated January 7, 2016, via UPS to the affected customers. The notification informed customers of the issue with the affected product; how to identify affected product; measures to be taken by the users; and that a MAQUET service representative will be contacting them to schedule servicing of the device to install a Support Frame Lock. Customers with questions were asked to contact their local MAQUET Representative or call MAQUET Customer Service at 1-888-627-8383 (Option 2, Option 3) Monday through Friday between 8:00 am and 5:00 pm EDT. Customers were also requested to fill out and return the attached response forms.

Details

Recall number
Z-1045-2016
Initiated
January 7, 2016
Posted by FDA
March 3, 2016
Terminated
May 25, 2017
Product code
FQO – Table, Operating-Room, Ac-Powered
Quantity
2 within US; 247 OUS
Distribution
Worldwide Distribution - US including AZ and MO. Foreign distribution to Australia, Austria, Bahrain, Belgium, Brazil, Canada, Denmark, Ecuador, Estonia, Finland, France, Germany, Hong Kong, India, Italy, Japan, Latvia, Malaysia, Netherlands, Norway, Poland, Portugal, Russia, Saudi Arabia, Serbia, South Korea, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom, and Venezuela.
Recalling firm
Maquet Cardiovascular Us Sales, Llc, Wayne, NJ

Official FDA recall record