Recall Z-1050-2015

Hospira Inc. · initiated June 11, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Plum A+ Infusion System with Hospira MedNet List 20792-04-25, Pole Clamp Assembly Part Number 820-95443-008. A cassette based multi-function infusion system. Each system includes a pumping module and an assortment of disposable IV sets, optional accessories, and an operator's manual. The pole clamp assembly is used to provide a mechanism to tighten and loosen the grip of the pole clamp to remove the infuser from the IV pole.

Codes / lots: List Number: 20792-04-25; Serial Numbers: 17737372 ; 17737375 ; 17737377 ; 17737378 ; 17737383 ; 17737385 ; 17737390 ; 17737394 ; 17737395 ; 17737396 ; 17737397 ; 17737398 ; 17737399 ; 17737400 ; 17737402 ; 17737405 ; 17737408 ; 17737409 ; 17737414 ; 17737416 ; 17737419 ; 17737420 ; 17737421 ; 17737422 ; 17737424 ; 17737425 ; 17737426 ; 17737429 ; 17737430 ; 17737431 ; 17737432 ; 17737433 ; 17737435 ; 17737438 ; 17737440 ; 17737442 ; 17737443 ; 17737444 ; 17737445 ; 17737446 ; 17737447 ; 17737448 ; 17737449 ; 17737450 ; 17737451 ; 17737452 ; 17737454 ; 17737456 ; 17737458 ; 17737459 ; 17737460 ; 17737462 ; 17737463 ; 17737464 ; 17737467 ; 17737470 ; 17737475 ; 17737476 ; 17737478 ; 17737479 ; 17737480 ; 17737483 ; 17737485 ; 17737486 ; 17737489 ; 17737490 ; 17737492 ; 17737498 ; 17737500 ; 17737501 ; 17737502 ; 17737506 ; 17737507 ; 17737508 ; 17737511 ; 17737514 ; 17737516 ; 17737518 ; 17737520 ; 17737521 ; 17737522 ; 17737524 ; 17737525 ; 17737526 ; 17737531 ; 17737533 ; 17737535 ; 17737539 ; 17737542 ; 17737545 ; 17737546 ; 17737550 ; 17737551 ; 17737556 ; 17737557 ; 17737558 ; 17737559 ; 17737560 ; 17737562 ; 17737564 ; 17737569 ; 17737575

Reason for recall

Plum A+ infusion system pole clamp assemblies were discovered to be cracked. If the knob on the pole clamp assembly is cracked, the infuser cannot be attached to the IV pole since the knob is used to tighten and loosen the grip of the pole clamp to the pole.

FDA-determined cause: Vendor change control

Action taken

Hospira contacted the the sole customer on June 11, 2014, to inform them of the issue of the potential impact to their Plum A+ infusers list 20792-04-25. Hospira requested that the customer to allow Hospira Field Service Representatives to inspect and replace the impact pole clamp assemblies. On June 12, 2013 Geneva Hospital approved Hospira's request to the inspection and replacement of the impacted pole clamp assemblies. On June 16, 2013 Hospira's Field Service personnel performed the on-site, visual inspection of the pole clamp assemblies and confirmed 15 units were cracked. Hospira determined to replace all 102 pole clamp assemblies that were received by Geneva Hospital. For further questions please call (800) 441- 4100.

Details

Recall number
Z-1050-2015
Initiated
June 11, 2014
Posted by FDA
February 4, 2015
Terminated
November 28, 2016
Product code
FRN – Pump, Infusion
Quantity
102 infusors
Distribution
US Distribution to New York.
Recalling firm
Hospira Inc., Lake Forest, IL

Official FDA recall record