Recall Z-1051-2015

Teleflex Medical · initiated January 12, 2015

Class I Terminated

Class I – dangerous or defective products that could cause serious health problems or death

Product

MAQUET, SERVO Humidifier 163, Rx Only, Inmed Mfg. Sdn Bhd. Catalog No: XKC01-06-8125-8, Model No. 01-06-8125-8. A heat and moisture condenser (artificial nose) is a device intended to be positioned over a tracheotomy (a surgically created opening in the throat) or tracheal tube (a tube inserted into the trachea) to warm and humidify gases breathed in by a patient.

Codes / lots: Lot numbers: 201413, 201414, 201415, 201417, 201419 and 201422.

Reason for recall

Defects: Cracks that may occur in the center of the corrugated flex tube of the 15 mm and 22 mm ISO female connector.

FDA-determined cause: Process control

Action taken

Consignees were notified by an Urgent - Field Safety Notice, dated 01/12/2015. The letter identified the affected devices, a description of the problem, and actions to be taken. Customers were asked to check their inventory for any affected product, to cease use and distribution, and immediately quarantine any affected product. Customers are to complete and return the Acknowledgement form and to their local MAQUET representative, who provide a return number. A credit will be issues upon receipt of the affected product. Customer should contact their local MAQUET representative at fieldactions@maquet.com with questions or for support this issue.

Details

Recall number
Z-1051-2015
Initiated
January 12, 2015
Posted by FDA
February 6, 2015
Terminated
May 10, 2016
Product code
BYD – Condenser, Heat And Moisture (Artificial Nose)
Quantity
100 units
Distribution
Distributed to FL, KS, MI and WV.
Recalling firm
Teleflex Medical, Durham, NC

Official FDA recall record