Recall Z-1052-2015

Zimmer, Inc. · initiated December 11, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Persona (TASP) Tibial Articular Surface Provisional Shim, all sizes and Lots Catalog Number: 42-5279-002-00 thru 42-5279-009-04. Product Usage: The Persona TASPs are sterilizable instruments Intended for multiple uses. The TASP components are used to aid the surgeon in assessing the thickness of the articular surface required. The TASP construct is utilized while conducting a trial range of motion. The TASP consists of 3 parts: a TASP bottom, a TASP shim, and a TASP top. The shims come in varying thicknesses to allow trialing for specific Articular Surface implant thicknesses.

Codes / lots: Persona (TASP) Tibial Articular Surface Provisional Shim, all sizes and Lots Catalog Number: 42-5279-002-00 thru 42-5279-009-04: 42-5279-002-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, CD, 10mm; 42-5279-002-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, CD, 12mm; 42-5279-002-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, CD, 14mm; 42-5279-004-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, EF, 10mm; 42-5279-004-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, EF, 12mm; 42-5279-004-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, EF, 14mm; 42-5279-006-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, GH, 10mm; 42-5279-006-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, GH, 12mm; 42-5279-006-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, GH, 14mm; 42-5279-008-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, J, 10mm; 42-5279-008-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, J, 12mm; 42-5279-008-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, J, 14mm; 42-5279-003-00 All Lots Persona TASP (Tibial A/S Provisional)…

Reason for recall

Potential for delay in surgery and/or ball bearings to be left in the wound. Repeated cycles of sonic cleaning can break down the ball bearing retention feature (swage) on the shim. To prevent this, TASP shims must be manually cleaned and inspected per product insert, 87-6203-991-22.

FDA-determined cause: Device Design

Action taken

Zimmer sent an URGENT MEDICAL DEVICE CORRECTION notifications letter dated December 11, 2014 to affected distributors via electronic mail, and all hospital Risk Managers and distributors with affected inventory via courier. The letter identified the affected product, problem, instructions for responding to the formal recall notification.and actions to be taken. For questions call the customer call center at 1-800-348-2759.

Details

Recall number
Z-1052-2015
Initiated
December 11, 2014
Posted by FDA
February 4, 2015
Terminated
October 23, 2015
Product code
OIY – Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer + Additive/Metal/Polymer + Additive
Quantity
67963
Distribution
Worldwide Distribution - USA Nationwide including DC, and PR and the countries of Australia, Austria, Belgium, Canada, Czech Republic, Dominican Republic, France, Germany, India, Iran, Israel, Italy, Japan, Luxembourg, Netherlands, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
Recalling firm
Zimmer, Inc., Warsaw, IN

Official FDA recall record