Recall Z-1053-2008

Draeger Medical, Inc. · initiated January 31, 2008

Terminated

Product

Draeger Globe Trotter IC, Catalog Number: MU20509, Draeger Medical, Inc., Telford, PA 18969

Codes / lots: Serial Numbers: QT05675, QT05676, ST05708, and ST05704

Reason for recall

Heating failure- The power board which controls the incubator heater may not regulate the temperature properly resulting in a high temperature alarm and a loss of temperature control within the patient compartment.

FDA-determined cause: Nonconforming Material/Component

Action taken

The firm notified consignees on 2/1/08, by an Important Safety Notice. The letter instructed users to remove any units from use until a Draeger Service Representative can replace the power board. A Service Representative will be in contact to schedule a service visit. For questions concerning this recall please contact 215-660-2418 or Anbu.arasu@draeger.com.

Details

Recall number
Z-1053-2008
Initiated
January 31, 2008
Posted by FDA
March 18, 2008
Terminated
May 8, 2008
Product code
FMZ – Incubator, Neonatal
Quantity
4 units
Distribution
Worldwide Distribution-- USA including states MO, OH, and PR, and countries of Australia, Canada, Columbia, Israel, UAE, France, China, Italy, Denmark, Switzerland, and UK.
Recalling firm
Draeger Medical, Inc., Telford, PA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1051-2008 Draeger TI500. Isolette Infant Incubator. Catalog Number: MU20505. Draeger Medical…Heating failure- The power board which controls the incubator heater may not regulate the temperature… 01/31/2008Terminated
Z-1052-2008 Draeger GT500. GlobeTrotter Transport System. Catalog Number: MU20508, Draeger Medical…Heating failure- The power board which controls the incubator heater may not regulate the temperature… 01/31/2008Terminated