Recall Z-1054-2025
Karl Storz Endoscopy · initiated December 19, 2024
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
KARL STORZ - ENDOSKOPE, REF: 27002KP, Pediatric Uretero-Renoscope 7.3 Fr., NON STERILE, CE 0123
Codes / lots: All Lots/UDI: 04048551231234
Reason for recall
Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.
FDA-determined cause: Labeling Change Control
Action taken
On December 19, 2024, Karl Storz, issued an "Urgent: Medical Device Correction" notification to affected consignees via UPS. In addition to informing consignees about the recall, Karl Storz asked consignees to take the following actions: 1. Immediately discard any previous revisions of the IFU. 2. Access the updated e-IFU at the following link: https://www.karlstorz.com/de/en/eifu.htm?q=&country=US 3. Pass on this Urgent Medical Device Correction to all users of the recalled products and all other persons who need to be aware within your organization. 4. If you have or may have distributed the products listed, please identify and promptly notify those recipients with a copy of this notice, response form and updated e-IFU. 5. Return the completed Customer Reply Form by Fax or E-Mail to the indicated contact below within 15 calendar days from the date of receipt. 6. Please report any incidents related to this issue to the manufacturer, dealer, or local representative and, if applicable, to the national competent authority, as this is important feedback. 7. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Details
- Recall number
- Z-1054-2025
- Initiated
- December 19, 2024
- Posted by FDA
- January 30, 2025
- Product code
- FGB – Ureteroscope And Accessories, Flexible/Rigid
- Quantity
- 39 units
- Distribution
- US Nationwide distribution.
- Recalling firm
- Karl Storz Endoscopy, El Segundo, CA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1051-2025 Class II | KARL STORZ - ENDOSKOPE, REF: 27003L, MICHEL Uretero-Renoscope, 9.5 Fr., 43cm, RxONLY…Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and… | 12/19/2024Open, Classified |
| Z-1052-2025 Class II | KARL STORZ - ENDOSKOPE, REF: 27030KA, Cysto-Urethroscope, NON STERILE, CEScope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and… | 12/19/2024Open, Classified |
| Z-1053-2025 Class II | KARL STORZ - ENDOSKOPE, REF: 27030KB, Pediatric Cysto-Urethroscope, NON STERILE, CEScope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and… | 12/19/2024Open, Classified |
| Z-1055-2025 Class II | KARL STORZ - ENDOSKOPE, REF: 27010L, Uretero-Renoscope, 7 Fr., 43cm, NON STERILE…Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and… | 12/19/2024Open, Classified |
| Z-1056-2025 Class II | KARL STORZ - ENDOSKOPE, REF: 27000L,Uretero-Renoscope, 7 Fr., 6, 43 cm, NON STERILE…Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and… | 12/19/2024Open, Classified |
| Z-1057-2025 Class II | KARL STORZ - ENDOSKOPE, REF: 27001L, Uretero-Renoscope, 8 Fr., 6, 43 cm, NON STERILE…Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and… | 12/19/2024Open, Classified |
| Z-1058-2025 Class II | KARL STORZ - ENDOSKOPE, REF: 27011L, Uretero-Renoscope, 8 Fr., 6, 43cm, NON STERILE…Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and… | 12/19/2024Open, Classified |
| Z-1059-2025 Class II | KARL STORZ - ENDOSKOPE, REF: 27002L, Uretero-Renoscope, 9.5 Fr., 6, 43 cm, NON STERILE…Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and… | 12/19/2024Open, Classified |
| Z-1060-2025 Class II | KARL STORZ - ENDOSKOPE, REF: 27010K, Ureteroscope, 7 Fr., 34 cm, NON STERILE, RxONLY, CE…Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and… | 12/19/2024Open, Classified |
| Z-1061-2025 Class II | KARL STORZ - ENDOSKOPE, REF: 27000K, Ureteroscope, 7 Fr., 6, 34 cm, NON STERILE, RxONLY…Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and… | 12/19/2024Open, Classified |
| Z-1062-2025 Class II | KARL STORZ - ENDOSKOPE, REF: 27001K, Ureteroscope, 9.5 Fr., 6, 34 cm, NON STERILE…Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and… | 12/19/2024Open, Classified |
| Z-1063-2025 Class II | KARL STORZ - ENDOSKOPE, REF: 27002K, Ureteroscope, 9.5 Fr., 6, 34 cm, NON STERILE…Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and… | 12/19/2024Open, Classified |
| Z-1064-2025 Class II | KARL STORZ - ENDOSKOPE, REF: 27011K, Ureteroscope, 8 Fr., 6, 34cm, CE 0123Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and… | 12/19/2024Open, Classified |