Recall Z-1064-2009

Abbott Laboratories · initiated November 17, 2008

Terminated

Product

CELL-Dyn Ruby Hematology Analyzer, List Number 08H67-01 The CELL-DYN Ruby is a multi-parameter, automated hematology analyzer designed for in vitro diagnostic use in clinical laboratories. The instrument has two modes of operation: the Autoloader mode that aspirates samples from closed collection tubes and the Open Tube mode that aspirates samples from open collection tubes.

Codes / lots: Serial No. 34001BG - 34965BG

Reason for recall

Carryover failures in software revisions 1.0ML and 2.0ML result in elevated PLT background count.

FDA-determined cause: Other

Action taken

Recall initiated on 11/17/2008. A Product Correction letter and customer reply form is sent to all currently active CELL-DYN Ruby customers. The customers are asked to keep the communication with their CELL-DYN Ruby System Operator's Manual and to note that the information listed in the table replaces the carryover specification provided in all applications of the manual. Questions from customers in the US should be directed to Customer Support at 1-877-4ABBOTT. Customers outside the US should contact their local hematology customer support representative.

Details

Recall number
Z-1064-2009
Initiated
November 17, 2008
Posted by FDA
April 10, 2009
Terminated
November 27, 2010
Product code
GKZ – Counter, Differential Cell
Quantity
899 units
Distribution
Worldwide Distribution
Recalling firm
Abbott Laboratories, Santa Clara, CA

Official FDA recall record