Recall Z-1065-2010

Hill-Rom, Inc. · initiated January 1, 2010

Terminated

Product

Hill-Rom TotalCare Bed System, Hill-Rom, Batesville, IN; Model P1900 and P1840. The product is used as an AC powered adjustable hospital bed.

Codes / lots: Serial numbers K285AM5539 through K310AM5864. (Serial numbers K310AM5857 , K296AM5635 , K296AM5629 , K295AM5623 , K306AM5779 , K306AM5777 , K303AM5747 , K303AM5741 , K300AM5677 , K300AM5675 , K299AM5654 , K299AM5651 , K302AM5718 , K301AM5715 , K301AM5712 , K301AM5710 , K301AM5709 , K301AM5707 , K301AM5705 , K301AM5704 , K301AM5702 , K301AM5700 , K301AM5696 , K301AM5695 , K300AM5693 , K300AM5690 , K300AM5686 , K300AM5682 , K299AM5660 , K299AM5653 , K296AM5647 , K296AM5643 , K299AM5656 , K299AM5652 , K302AM5725 , K302AM5723 , K302AM5719 , K301AM5714 , K301AM5698 , K300AM5688 , K299AM5663 , K299AM5658 , K302AM5735 , K302AM5733 , K302AM5731 , K302AM5729 , K302AM5726 , K302AM5724 , K302AM5722 , K302AM5720 , K302AM5716 , K301AM5713 , K301AM5708 , K301AM5706 , K300AM5691 , K294AM5594 , K303AM5761 , K303AM5748 , K303AM5739 , K303AM5760 , K303AM5759 , K303AM5758 , K303AM5757 , K303AM5756 , K303AM5755 , K303AM5754 , K303AM5753 , K303AM5752 , K303AM5751 , K303AM5750 , K303AM5749 , K303AM5738 , K296AM5648 , K296AM5645 , K296AM5642 , K306AM5780 , K306AM5776 , K303AM5740 , K302AM5737 , K302AM5734 , K302AM5732 , K302AM5730 , K302AM5728 , K302AM5721 , K300AM5678 , K300AM5676 , K300AM5674…

Reason for recall

The controls for nurse call, bed movement and entertainment functions on the intermediate rail may fail to function.

FDA-determined cause: Nonconforming Material/Component

Action taken

The firm sent a letter to each consignee dated January 4, 2010 advising them that a firm representative will visit the facility to make corrections. Hospitals were requested to take all of the affected beds out of service, if possible, until they are corrected, or to inspect and remove those with a problem out of service until they are corrected if all cannot be removed from service. If you have any questions concerning the recall, please contact Hill-Rom Technical Support at 1-800-445-3720 (U.S.).

Details

Recall number
Z-1065-2010
Initiated
January 1, 2010
Posted by FDA
March 11, 2010
Terminated
February 17, 2012
Product code
FNL – Bed, Ac-Powered Adjustable Hospital
Quantity
185
Distribution
Worldwide Distribution -- United States, Australia, Canada, France, Germany, Mexico, United Arab Emirates and United Kingdom.
Recalling firm
Hill-Rom, Inc., Batesville, IN