Recall Z-1065-2010
Hill-Rom, Inc. · initiated January 1, 2010
Product
Hill-Rom TotalCare Bed System, Hill-Rom, Batesville, IN; Model P1900 and P1840. The product is used as an AC powered adjustable hospital bed.
Codes / lots: Serial numbers K285AM5539 through K310AM5864. (Serial numbers K310AM5857 , K296AM5635 , K296AM5629 , K295AM5623 , K306AM5779 , K306AM5777 , K303AM5747 , K303AM5741 , K300AM5677 , K300AM5675 , K299AM5654 , K299AM5651 , K302AM5718 , K301AM5715 , K301AM5712 , K301AM5710 , K301AM5709 , K301AM5707 , K301AM5705 , K301AM5704 , K301AM5702 , K301AM5700 , K301AM5696 , K301AM5695 , K300AM5693 , K300AM5690 , K300AM5686 , K300AM5682 , K299AM5660 , K299AM5653 , K296AM5647 , K296AM5643 , K299AM5656 , K299AM5652 , K302AM5725 , K302AM5723 , K302AM5719 , K301AM5714 , K301AM5698 , K300AM5688 , K299AM5663 , K299AM5658 , K302AM5735 , K302AM5733 , K302AM5731 , K302AM5729 , K302AM5726 , K302AM5724 , K302AM5722 , K302AM5720 , K302AM5716 , K301AM5713 , K301AM5708 , K301AM5706 , K300AM5691 , K294AM5594 , K303AM5761 , K303AM5748 , K303AM5739 , K303AM5760 , K303AM5759 , K303AM5758 , K303AM5757 , K303AM5756 , K303AM5755 , K303AM5754 , K303AM5753 , K303AM5752 , K303AM5751 , K303AM5750 , K303AM5749 , K303AM5738 , K296AM5648 , K296AM5645 , K296AM5642 , K306AM5780 , K306AM5776 , K303AM5740 , K302AM5737 , K302AM5734 , K302AM5732 , K302AM5730 , K302AM5728 , K302AM5721 , K300AM5678 , K300AM5676 , K300AM5674…
Reason for recall
The controls for nurse call, bed movement and entertainment functions on the intermediate rail may fail to function.
FDA-determined cause: Nonconforming Material/Component
Action taken
The firm sent a letter to each consignee dated January 4, 2010 advising them that a firm representative will visit the facility to make corrections. Hospitals were requested to take all of the affected beds out of service, if possible, until they are corrected, or to inspect and remove those with a problem out of service until they are corrected if all cannot be removed from service. If you have any questions concerning the recall, please contact Hill-Rom Technical Support at 1-800-445-3720 (U.S.).
Details
- Recall number
- Z-1065-2010
- Initiated
- January 1, 2010
- Posted by FDA
- March 11, 2010
- Terminated
- February 17, 2012
- Product code
- FNL – Bed, Ac-Powered Adjustable Hospital
- Quantity
- 185
- Distribution
- Worldwide Distribution -- United States, Australia, Canada, France, Germany, Mexico, United Arab Emirates and United Kingdom.
- Recalling firm
- Hill-Rom, Inc., Batesville, IN