Recall Z-1068-2011
Ethicon Endo-Surgery Inc · initiated March 27, 2009
Product
ENDOPATH ETS-FLEX45 Articulating Endoscopic Linear Cutters, Product Codes: ATS45; LONG45A; 6TB45; ATG45; ATB45; ATW45; ATS45NK; and NAW45. Intended for transection, resection, and/or creation of anastomoses.
Codes / lots: The recalled Lot Numbers are: E4MJ47, E4MF6W, E4MK84, E4MM23, E4MM64, E4MP4A, E4MR15, E4MT07, E4MT4Y, F4MU2R, F4MY91, E4MG53, E4MG5K, E4MG5L, E4MN39, E4MG89, E4MG9M, E4MH2L, E4ML6J, E4MG77, E4MG8A, E4MH18, E4MJ3K, F4MX8C, and F4N799.
Reason for recall
The firm initiated this second subsequent recall because the affected devices can remain closed and clamped on tissue during surgical procedures due to the fact that the mechanism which connects the articulation joint to the device shaft is insufficient.
FDA-determined cause: Other
Action taken
Ethicon Endo-Surgery issued "Urgent Expanded Device Recall Event 1981" notification letters dated April 7, 2009, to their customers. They provided a list of all Affected Product Codes of the ETS FLEX45 Articulating Endoscopic Linear Cutters (including the Product Code: ATS45 from the 1st recall) and Affected Flex Tray Codes subject to recall. The notification document reminds the Operating Room and Materials Management customers of the problem which necessitated the previous recall (Z-0361-2011). It goes on to instruct their customers to immediately discontinue use of the affected products. They are to complete and fax back the Business Reply Form, Affected product is to be returned with a copy of the completed Business Reply Form to Stericycle, Replacement product should be received within 30-60 days after return, Customers can contact Stericycle concerning this recall at 1-888-365-4680, reference Event 1981.
Details
- Recall number
- Z-1068-2011
- Initiated
- March 27, 2009
- Posted by FDA
- February 15, 2011
- Terminated
- February 15, 2013
- Product code
- GEI – Electrosurgical, Cutting & Coagulation & Accessories
- Quantity
- 29,403 units
- Distribution
- Nationwide Distribution -- USA including the following states, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MO, MS, MT, NC, NE, NH, NJ, MN, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, and Puerto Rico.
- Recalling firm
- Ethicon Endo-Surgery Inc, Cincinnati, OH