Recall Z-1068-2017

The Binding Site Group, Ltd. · initiated September 11, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Human lgG4 Subclass kit for use on the Beckman Coulter AU" series This kit is intended for qualifying human IgG subclasses 1,2,3 and 4 immunoglobulins in serum on the Beckman Coulter AU series. Meaurement of these immunoglobulins ais in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents Product Code LK009.AU

Codes / lots: All lots.

Reason for recall

Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.

FDA-determined cause: Nonconforming Material/Component

Action taken

Binding Site sent an Important Information letter dated September 30, 2014, Customers were instructed to retain the notification for records and additional instructions with the firm stating they were updating the instructions for use. Customers with questions were advised to contact their Binding Site representative or technical.service@bindingsite.co.uk.

Details

Recall number
Z-1068-2017
Initiated
September 11, 2014
Posted by FDA
January 23, 2017
Terminated
May 3, 2017
Product code
CFN – Method, Nephelometric, Immunoglobulins (G, A, M)
Distribution
Worldwide Distribution - US including n FL, OH, TN, MN, NW, MI, CA. and Internationally to Canada, France, UK, Spain, Germany, Czech Republic, Belgium, Italy, Australia, Brazil, Chile, China, Costa Rica, Kuwait, New Zealand, Oman, Pakistan, Portugal, Sweded, Taiwan, and Turkey.
Recalling firm
The Binding Site Group, Ltd., Birmingham

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1067-2017 Class IIHuman Subclass Kit for use on the Beckman IMMAGE"/IMMAGE"800 Analyser Product Code…Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and… 09/11/2014Terminated
Z-1069-2017 Class IIHuman lgG4 Subclass Liquid Reagent Kits For use on the Roche Cobas"6000 This kit is…Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and… 09/11/2014Terminated
Z-1070-2017 Class IIHuman lgG4 Subclass Liquid Reagent Kits Human lgG1& lgG2 Antisera, lgG3 & 4 Latex…Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and… 09/11/2014Terminated
Z-1071-2017 Class IIHuman lgG4 Subclass Liquid Reagent Kits for use on the Siemens BN"II Analyser Product…Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and… 09/11/2014Terminated
Z-1072-2017 Class IIHuman lgG Subclass Liquid Reagent Kits for use on the Siemens BN"IJ Analyser Product…Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and… 09/11/2014Terminated
Z-1073-2017 Class IIMininephThl Human lgG4 Kit Product Code: ZK009.LR This kit is intended for qualifying…Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and… 09/11/2014Terminated