Recall Z-1070-2012
Benlan, Inc. · initiated December 12, 2011
Product
Children's Medical Ventures 1 ml Amber Oral Syringe. The product is a Liquid Medication Dispenser.
Codes / lots: Catalog number 1047080 - lot numbers 076192, 076404, 076904, 077695, 078697, 079540 and 079686.
Reason for recall
The tip caps may come loose from the syringe, potentially resulting in leakage of medication from the syringe.
FDA-determined cause: Process control
Action taken
The firm initiated their recall of the specified lots of product in November, 2011 by sending a "Customer/Distributor Medical Device Correction" letter to their consignee. The letter requested they examine their inventory, quarantine any affected product and complete a verification and response form to be returned to the firm. Questions were directed to (905) 829-5004 ext 233.
Details
- Recall number
- Z-1070-2012
- Initiated
- December 12, 2011
- Posted by FDA
- February 21, 2012
- Terminated
- March 19, 2013
- Product code
- KYX – Dispenser, Liquid Medication
- Quantity
- 143,250
- Distribution
- Nationwide Distribution, USA - including only the state of Pennsylvania
- Recalling firm
- Benlan, Inc., Oakville