Recall Z-1079-2014

Zimmer, Inc. · initiated October 17, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Trabecular Metal" Reverse Shoulder System

Codes / lots: 61025613, 61009830, 61038635, 61031385, 61021507, 61021506, 61016907, 61009834, 61006493, 61006492, 60999239, 60993837, 60993836, 60987295, 60982839, 60982838, 60970974, 60970973, 60950253, 60964425, 60946627, 60970946, 60964426, 60943695, 60940448, 60950251, 60946624, 60940449, 60927405, 60934492, 60900542, 60927406, 60919029, 60914617, 60919028, 60900543, 60893180, 60893181, 60906211, 60887745, 60887744, 60875337, 60875336, 60868040, 60873954, 60868041, 60867386, 60867385, 60856882, 60856880, 60855532, 60852815, 60855529, 60846158, 60846157, 60840619, 60834839, 60806512, 60824952, 60820211, 60814537, 60785523, 60795705, 60792083, 60761203, 60752933, 60781096, 60776583, 60765951, 60768804, 60757206, 60747093, 60741590, 60747092, 60741589, 60738183, 60727960, 60738184, 60730072, 60730073, 60723302, 60723277, 60718989, 60712568, 60683321, 60675599, 60669310, 60674468, 105533, 60605155, 105588, 60577323, 60580942, 60572796, 60564704, 60572798, 60568536, 60568535, 60578677, 60568468, 60568467, 60568466, 60564623, 60564639, 60564628, 60564641, 60564638, 60564636, 60568470, 60564642, 60564635, 60564625, 60564621, and 60564633.

Reason for recall

As a result of complaints for disassociation of the TM Reverse Glenosphere from the Baseplate, Zimmer updated the related surgical technique and released a video to provide additional instructions to users on how to properly assemble the Glenosphere to the Baseplate.

FDA-determined cause: Device Design

Action taken

In November and December 2012 Sales staff were provided updated surgical technique and associated video to share with current customers, surgeons, and risk managers. October 17, 2013 Distributors ( via E-mail),and Risk Managers and Surgeons (via traceable courier) were sent written Urgent Recall notifications of these labeling changes. Current customers are identified through an invoice search and distribution history for the affected devices for the past 12 months, since 9/1/12.

Details

Recall number
Z-1079-2014
Initiated
October 17, 2013
Posted by FDA
February 24, 2014
Terminated
September 12, 2014
Product code
KWT – Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented
Quantity
73,002 units
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Germany, India, Malaysia, China, Singapore, Korea, Taiwan, and Brazil.
Recalling firm
Zimmer, Inc., Warsaw, IN

Official FDA recall record