Recall Z-1084-2022

Beckman Coulter Inc. · initiated October 21, 2019

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

MicroScan WalkAway-96 plus Instrument REF B1018-284

Codes / lots: Catalog Number: B1018-284 UDI-DI Code: 15099590658687 Serial Numbers: 39610007 39610008 39610009 39610010 39610011 39610012 39610013

Reason for recall

WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere

FDA-determined cause: Mixed-up of materials/components

Action taken

During the week of 10/21/2019, Beckman Coulter sent an "URGENT MEDICAL DEVICE RECALL" Letter via e-mail and postal mail informing customers that some WalkAway instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1A (ampere) current rating rather than the correct current rating at 10A. Customers are instructed to: If the component fails and and the WalkAway suffers a power loss, user should: --Contact the Customer Support Center. If in the United States call 1-800-677-7226. If outside the United States, contact their local Beckman Coulter Representive. --Determine if the customer knows how long the WalkAway has been without power. If the time is unknown or greater than 2 hours, then panels should be reset, or an alternate test method should be used. -- If within 2 hours window the instrument will maintain the environmental conditions. Depending on the panel type, the user has the option to read the panels manually, on an autoSCAN-4 (if available) or reset the panels. - If the component remains functional, a Beckman Coulter Field Service Representative will contact the user to schedule an onsite visit. During the visit the representative will inspect the suspect instrument and if the incorrect component is found, will replace the component with the correct one rated at 10A. -Complete and return Customer Response Form to Beckman Coulter, Inc. 1584 Enterprise Blvd. West Sacramento, CA 95691 Attn: Quality Systems & Compliance Fax number: 916-374-2119 Email address: MicrobiologyCustomersHCUS@beckman.com For questions, contact: -via website, http://www.beckmancoulter.com/customersupport/support -via phone, call 1-800-677-7226 in the United States, -If outside the United States, contact the local Beckman Coulter Representative.

Details

Recall number
Z-1084-2022
Initiated
October 21, 2019
Product code
LRG – Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems
Quantity
7 systems
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of Alaska, Arkansas, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Missouri, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Pennsylvania, Tennessee, Texas, Utah, Vermont, Virginia, and Washington. The…
Recalling firm
Beckman Coulter Inc., Brea, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1083-2022 Class IIMicroScan WalkAway-40 plus Instrument REF B1018-283WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated… 10/21/2019Open, Classified
Z-1085-2022 Class IIDxM 1040 MicroScan WalkAway Instrument REF B1018-440WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated… 10/21/2019Open, Classified
Z-1086-2022 Class IIDxM 1096 MicroScan WalkAway Instrument REF B1018-496WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated… 10/21/2019Open, Classified
Z-1087-2022 Class IIMicroScan WalkAway-40 plus Reconditioned Instrument REF B1018-340RWalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated… 10/21/2019Open, Classified
Z-1088-2022 Class IIMicroScan WalkAway-96 plus Reconditioned Instrument REF B1018-396RWalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated… 10/21/2019Open, Classified