Stryker, Inc. · initiated December 6, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Stryker Neurovascular, Trevo Pro 18 Microcatheter, REF 90238 0.021in X150cm

Codes / lots: UDI: (01)00815742002386/Lot Number(s): 0000228865 0000370674

Reason for recall

The microcatheter IFU contains an intended use that has not been evaluated by the FDA.

FDA-determined cause: Under Investigation by firm

Action taken

On December 22, 2023, Stryker Neurovascular issued a "Urgent: Medical Device Recall" notification to affected consignees. Stryker asked consignees to take the following actions: Immediately check your internal inventory for affected devices. 2. The affected items may continue to be used. 3. When using impacted lot(s) of Trevo Trak 21 or Trevo Pro 18 microcatheter, please follow the IFU attached in this communication instead of the one that originally came with the device. 4. Circulate this Customer Notification internally to all interested/affected parties. 5. Maintain awareness of this communication internally until all required actions have been completed within your facility. 6. Inform Stryker if any of the subject devices have been distributed to other organizations. If yes, provide contact details so that Stryker can inform the recipients appropriately. 7. Please inform Stryker of any adverse events concerning the use of the subject devices. (US Customer Service: 1-855-91-NEURO (1-855-916-3876), Email: NVCustomerCare@stryker.com) 8. Complete the attached customer response form. It may be that you no longer have any physical inventory on site. Completing this form will allow us to update our records and will also negate the need for us to send any further unnecessary communications on this matter. 9. Email the completed form to nvfieldactions@stryker.com

Details

Recall number
Z-1084-2024
Initiated
December 6, 2023
Posted by FDA
February 14, 2024
Product code
DQO – Catheter, Intravascular, Diagnostic
Quantity
11 units
Distribution
US Nationwide distribution in the states of AL, FL, GA, HI, IL, MI, NC, NH, OH, PA, TX, VA.
Recalling firm
Stryker, Inc., Salt Lake City, UT

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1085-2024 Class IIStryker Neurovascular, Trevo Trak 21 Microcatheter, REF 90338, 0.021in X162cmThe microcatheter IFU contains an intended use that has not been evaluated by the FDA. 12/06/2023Open, Classified