Recall Z-1087-2026

Medline Industries, LP · initiated December 16, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Medline medical procedure convenience kits labeled as: 1) WCH BARI-GASTRECTOMY CDS, REF CDS860237AF; 2) ROBOTIC PROCEDURE, REF CDS982128Q; 3) CUH THORACIC, REF CDS984338M; 4) THORACIC PROCEDURE, REF CDS984640I; 5) DAVINCI ROBOTIC PACK-LF, REF DYNJ0525321O; 6) LAP CHOLE PACK, REF DYNJ57640F; 7) GYN/PROSTATE ROBOTIC PACK, REF DYNJ62594D; 8) RR-DAVINCI PROSTATE ACCESSORY, REF DYNJ63392D; 9) LAPAROSCOPY PACK, REF DYNJ68187B; 10) LAPAROSCOPY PACK, REF DYNJ82368A; 11) ROBOTIC KIDNEY PACK, REF DYNJ85261; 12) DOWNTOWN ROBOTICS, REF DYNJ86178B; 13) DAVINCI LITHOTOMY PACK, REF DYNJ87610; 14) ROBOTIC WHIPPLE PACK, REF DYNJ88433; 15) PROSTATE DAVINCI, REF DYNJ900127G; 16) GENERAL LAPAROSCOPY, REF DYNJ901847M; 17) LAG TLH, REF DYNJ901890N; 18) LITHOTOMY ROBOTIC CDS, REF DYNJ903948S; 19) LAP, REF DYNJ904576B; 20) ROBOTIC, REF DYNJ905147B; 21) ROBOTICS GYN, REF DYNJ905824M; 22) WH DAVINCI TLH PROCEDURE, REF DYNJ905920B; 23) CHN OR BASIC ROBOTIC, REF DYNJ905982I; 24) KIT ROBOTIC RFD, REF DYNJ906311F; 25) ROBOTIC PROSTATE, REF DYNJ906381G; 26) DAVINCI, REF DYNJ907392C; 27) CHIASSON GASTRIC SLEEVE KIT, REF DYNJ907405C; 28) GYN, REF DYNJ908504B; 29) UROLOGY ROBOTIC, REF DYNJ908588I; 30) THORACIC ROBOTICS, REF DYNJ908777A; 31) GENERAL ROBOT, REF DYNJ908969A; 32) KIT ROBOTIC, REF DYNJ909870B; 33) ROBOTIC THORACOSCOPY, REF DYNJ910924.

Codes / lots: REF CDS860237AF: UDI/DI 10198459232626 (EA) 40198459232627 (CS), Lot Number 25FBK870; REF CDS860237AF: UDI/DI 10198459232626 (EA) 40198459232627 (CS), Lot Number 25EBD110; REF CDS982128Q: UDI/DI 10198459008245 (EA) 40198459008246 (CS), Lot Number 25EBC253; REF CDS982128Q: UDI/DI 10198459008245 (EA) 40198459008246 (CS), Lot Number 25DBG393; REF CDS984338M: UDI/DI 10198459148101 (EA) 40198459148102 (CS), Lot Number 25FBQ303; REF CDS984338M: UDI/DI 10198459148101 (EA) 40198459148102 (CS), Lot Number 25EBD631; REF CDS984338M: UDI/DI 10198459148101 (EA) 40198459148102 (CS), Lot Number 25DBM322; REF CDS984640I: UDI/DI 10198459318955 (EA) 40198459318956 (CS), Lot Number 25DBT807; REF DYNJ0525321O: UDI/DI 10198459236907 (EA) 40198459236908 (CS), Lot Number 25DMG099; REF DYNJ0525321O: UDI/DI 10198459236907 (EA) 40198459236908 (CS), Lot Number 25CMJ006; REF DYNJ57640F: UDI/DI 10198459011115 (EA) 40198459011116 (CS), Lot Number 25FDB347; REF DYNJ62594D: UDI/DI 10198459049767 (EA) 40198459049768 (CS), Lot Number 25EBK539; REF DYNJ62594D: UDI/DI 10198459049767 (EA) 40198459049768 (CS), Lot Number 25DBD145; REF DYNJ63392D: UDI/DI 10198459164507 (EA) 40198459164508 (CS), Lot Number 25BLA967…

Reason for recall

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

FDA-determined cause: Under Investigation by firm

Action taken

Medline Industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/16/2025 via email and USPS first class mail. The notice explained the issue, potential risk, and requested the following: "1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2.Please use the link and the information below to complete your response form. Please list the quantityof affected product you have in inventory on the form. Even if you do not have any affected productin inventory, please complete and submit the form. Website link: https://recalls.medline.comRecall Reference #: R-25-145-FGX5 Recall Code: 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4.If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com."

Details

Recall number
Z-1087-2026
Initiated
December 16, 2025
Posted by FDA
January 14, 2026
Product code
LRO – General Surgery Tray
Quantity
4536 units
Distribution
US Nationwide distribution.
Recalling firm
Medline Industries, LP, Northfield, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1085-2026 Class IIMedline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF…Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024… 12/16/2025Open, Classified
Z-1086-2026 Class IIMedline medical procedure convenience kits labeled as: 1) LAPAROSCOPY CDS-LF, REF…Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024… 12/16/2025Open, Classified
Z-1088-2026 Class IIMedline medical procedure convenience kits labeled as: 1) ROBOTIC, REF CDS980523M; 2)…Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024… 12/16/2025Open, Classified
Z-1089-2026 Class IIMedline medical procedure convenience kits labeled as: 1) LAVH PROCEDURE, REF…Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024… 12/16/2025Open, Classified
Z-1090-2026 Class IIMedline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE…Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024… 12/16/2025Open, Classified