Recall Z-1096-2025

Ethicon, LLC · initiated December 20, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

PROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, and neurosurgical procedures.

Codes / lots: REF/UDI-DI-GTIN/Lot(expiration): EH7585H/10705031097254/103DQR(8/31/2029); 8706H/10705031019430/103E22(8/31/2029), 103EMJ(8/31/2029); 8833H/10705031021037/103HC4(8/31/2029); 8963H/10705031091931/103E8Z(8/31/2029); 8935H/10705031022089/103HBP(8/31/2029)

Reason for recall

Sutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection.

FDA-determined cause: Process control

Action taken

On 12/20/2024, Johnson & Johnson Med Tech mailed recall notices to Operating Room Supervisors, Materials Management Personnel, Chief of Surgery, and distributors who were asked to do the following: 1) Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). 2) Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 3) Contact Sedgwick at 888-548-8526 to arrange affected product return. 4) Keep this notice visibly posted for awareness until all product subject to this recall has been returned to Sedgwick. 5) Complete and return the Business Reply Form via email to ethicon3703@sedgwick.com If you have additional questions regarding this communication contact the firm's Resource Department at 1-877-ETHICON (1-877-384-4266) open Monday through Friday, 8:00 AM to 5:00 PM ET.

Details

Recall number
Z-1096-2025
Initiated
December 20, 2024
Posted by FDA
February 4, 2025
Product code
GAW – Suture, Nonabsorbable, Synthetic, Polypropylene
Quantity
15,948
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, LA MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and the countries of AT, AU, BE, CA, CH, CN, CZ, DE, DK, FI, FR, GB, HU, IE, JP, LU, NL, NO, PL, SE, SG, ZA.
Recalling firm
Ethicon, LLC, San Lorenzo

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1097-2025 Class IIPERMA-HAND Silk Suture, REF: K872H, W723H, 623H, K833H; Perma-hand Silk Suture is…Sutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine… 12/20/2024Open, Classified
Z-1098-2025 Class IIETHIBOND EXCEL Polyester Suture, REF: 6664H, X425H, X872H, EH7491H. Not sold in USSutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine… 12/20/2024Open, Classified