Recall Z-1102-2017

PerkinElmer Health Sciences, Inc. · initiated December 21, 2016

Class III Terminated

Class III – products unlikely to cause any adverse health reaction but that violate FDA labeling or manufacturing laws

Product

DELFIA PC with V3.0 Software, Fluorometer, for clinical use Product Number: 1235-8100 Product Usage: The Wallac 1235 AutoDELFIA automatic immunoassay system is designed to automatically perform assays using the DELFIA technology. DELFIA technology is based on the proven and widely used method of time resolved fluorometry

Codes / lots: Batch Code US: 18RXWD2, 18TYWD2 OUS: B901XD2,18WYWD2, 18RXWD2

Reason for recall

AutoDELFIA System with Version 3.0 Software: Potential plate loading failures

FDA-determined cause: Software design

Action taken

PerkinElmer issued letter on 12/21/16 to customers with AutoDELFIA Systems running version 3.0 Manager Software (revisons 1 and 2). The issue cannot be corrected by the customer therefore it is recommended the AutoDELFIA System be operated in the default English language variant until the AutoDELFIA Manager Software can be updated. The AutoDELFIA System with version 3.0 Manager Software performs as intended when the selected language variant is English. Customers are recommended to implement a software update to be deployed by PerkinElmer to resolve the issue. AutoDELFIA Manager Software version 3.0 revision 1 or 2 will be updated through the correction to revision 3 or the latest released revision

Details

Recall number
Z-1102-2017
Initiated
December 21, 2016
Terminated
October 3, 2018
Product code
KHO – Fluorometer, For Clinical Use
Quantity
1 US ; 3 OUS
Distribution
Worldwide Distribution - US Nationwide in the state of MN and countries of China, India, Philippines, France, South Korea, Vietnam, Taiwan, Morocco, Kazachstan, UK, Spain, and Israel,
Recalling firm
PerkinElmer Health Sciences, Inc., Waltham, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1100-2017 Class IIIAutoDELFIA Plate Processor , Fluorometer, for clinical use Product Number: 1235-5110…AutoDELFIA System with Version 3.0 Software: Potential plate loading failures 12/21/2016Terminated
Z-1101-2017 Class IIIProgram Update Package AutoDELFIA 3.0, Fluorometer, for clinical use Product Number:…AutoDELFIA System with Version 3.0 Software: Potential plate loading failures 12/21/2016Terminated