Recall Z-1102-2019

Abbott Laboratories, Inc · initiated July 17, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Abbott ARCHITECT c4000 Processing Module, List No. 02P24 - Product Usage: The Abbott ARCHITECT System is intended for In Vitro diagnostic use only. The Abbott ARCHITECT Clinical Chemistry Systems are designed to perform automated chemistry tests, utilizing photometry and potentiometric technology. The ARCHITECT c4000, c8000, c16000 sample pipettor assemblies include a fluid sense/pressure monitoring system that helps identify errors in aspiration.

Codes / lots: All serial numbers

Reason for recall

There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.

FDA-determined cause: Software design

Action taken

Recall letters were issued in the U.S. 7/17/2018 via FedEx overnight mail notifying the consignee of the issue, patient impact, and necessary actions to take. Globally, primary methods of issuance were via mail, telephone, and email.

Details

Recall number
Z-1102-2019
Initiated
July 17, 2018
Terminated
September 23, 2020
Product code
JJE – Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Quantity
3,675 instruments
Distribution
Worldwide distribution - US Nationwide and countries Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Belarus, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Burkina FASO, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Curacao, Cyprus, Czech Republic, Democratic Rep. of Congo, Denmark…
Recalling firm
Abbott Laboratories, Inc, Irving, TX

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1103-2019 Class IIAbbott ARCHITECT c8000 Processing Module, List No. 01G06 - Product Usage: The Abbott…There is a potential to generate incorrect results on the instrument if particular error codes are not… 07/17/2018Terminated
Z-1104-2019 Class IIAbbott ARCHITECT c16000 Processing Module, List No. 03L77 - Product Usage: The Abbott…There is a potential to generate incorrect results on the instrument if particular error codes are not… 07/17/2018Terminated