Recall Z-1108-2008

Beckman Coulter Inc · initiated July 19, 2007

Terminated

Product

UniCel Dxl Access Immunoassay Systems Reaction Vessels for Use with UniCel Dxl Access Immunoassay Systems, Part Number : 386167, Beckman Coulter, Inc, Brea CA 92822

Codes / lots: Part Number: 386167; Lot Code: 1201291 & 1201431

Reason for recall

Erroneous Results - UniCel Dxl Access Immunoassay Systems Reaction Vessels have been found to be defective and there is a potential for an increase in signal leading to an erroneous result.

FDA-determined cause: Process design

Action taken

A Product Corrective Action (PCA) letter was mailed on July 19, 2007, to the Customers that received UniCel Dxl Access lmmunoassay System Reaction Vessels lot 1201291 and 1201431. The customers were informed that a small percentage of the reaction vessels (Lot #s: 1201291and 1201431 ) have been found to be defective and there is a potential for an increase in signal leading to an erroneous result. The customers were instructed to discontinue use and discard these lots. A response form was included.

Details

Recall number
Z-1108-2008
Initiated
July 19, 2007
Posted by FDA
March 19, 2008
Terminated
May 4, 2012
Product code
MMI – Immunoassay Method, Troponin Subunit
Quantity
264 units in the US; 9 units in Canada
Distribution
Nationwide & Canada
Recalling firm
Beckman Coulter Inc, Brea, CA