Recall Z-1108-2008
Beckman Coulter Inc · initiated July 19, 2007
Product
UniCel Dxl Access Immunoassay Systems Reaction Vessels for Use with UniCel Dxl Access Immunoassay Systems, Part Number : 386167, Beckman Coulter, Inc, Brea CA 92822
Codes / lots: Part Number: 386167; Lot Code: 1201291 & 1201431
Reason for recall
Erroneous Results - UniCel Dxl Access Immunoassay Systems Reaction Vessels have been found to be defective and there is a potential for an increase in signal leading to an erroneous result.
FDA-determined cause: Process design
Action taken
A Product Corrective Action (PCA) letter was mailed on July 19, 2007, to the Customers that received UniCel Dxl Access lmmunoassay System Reaction Vessels lot 1201291 and 1201431. The customers were informed that a small percentage of the reaction vessels (Lot #s: 1201291and 1201431 ) have been found to be defective and there is a potential for an increase in signal leading to an erroneous result. The customers were instructed to discontinue use and discard these lots. A response form was included.
Details
- Recall number
- Z-1108-2008
- Initiated
- July 19, 2007
- Posted by FDA
- March 19, 2008
- Terminated
- May 4, 2012
- Product code
- MMI – Immunoassay Method, Troponin Subunit
- Quantity
- 264 units in the US; 9 units in Canada
- Distribution
- Nationwide & Canada
- Recalling firm
- Beckman Coulter Inc, Brea, CA