Recall Z-1114-2008

Pointe Scientific, Inc. · initiated December 14, 2007

Terminated

Product

Pointe Scientific Liquid ALT (SGPT) reagent set for the quantitative determination of alanine aminotransferase in serum, 610 ml, Catalog no. HA926-610, in vitro diagnostic, Pointe Scientific, Inc.; Canton, MI 48188

Codes / lots: Lots 628302-004 and 704601-113.

Reason for recall

Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms.

FDA-determined cause: Process control

Action taken

Pointe Scientific notified consignees via Urgent Medical Device Recall letters dated 12/14/07, instructed to dispose of the product and to send back a reply form to request replacement product.

Details

Recall number
Z-1114-2008
Initiated
December 14, 2007
Posted by FDA
May 22, 2008
Terminated
September 15, 2008
Product code
CKA – Nadh Oxidation/Nad Reduction, Alt/Sgpt
Quantity
955 for all products.
Distribution
Worldwide Distribution - USA, Bolivia, Canada, Chile, China, Cyprus, Greece, Indonesia, Israel, Korea, Mexico, Pakistan, Panama, Peru, Philippines, Republic of Yemen, Russia, South Africa, Trinidad, Ukraine and West Indies.
Recalling firm
Pointe Scientific, Inc., Canton, MI

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1110-2008 Pointe Scientific Liquid ALT (SGPT) reagent set for the quantitative determination of…Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms. 12/14/2007Terminated
Z-1111-2008 Pointe Scientific Liquid ALT (SGPT) reagent set for the quantitative determination of…Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms. 12/14/2007Terminated
Z-1112-2008 Pointe Scientific Liquid ALT (SGPT) reagent set for the quantitative determination of…Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms. 12/14/2007Terminated
Z-1113-2008 Pointe Scientific Liquid ALT (SGPT) reagent set for the quantitative determination of…Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms. 12/14/2007Terminated
Z-1115-2008 Pointe Scientific Liquid ALT (SGPT) reagent set for the quantitative determination of…Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms. 12/14/2007Terminated
Z-1116-2008 Pointe Scientific Liquid ALT (SGPT) reagent set for the quantitative determination of…Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms. 12/14/2007Terminated
Z-1117-2008 Pointe Scientific Liquid ALT (SGPT) reagent set for the quantitative determination of…Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms. 12/14/2007Terminated