Recall Z-1114-2013
Roche Diagnostics Operations, Inc. · initiated March 6, 2013
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the CoagCHEK XS Pro Meters to quantitatively determines prothrombin time ("PT").
Codes / lots: catalog number 05530199160
Reason for recall
Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards. Errors have been confirmed with the EAN 13 barcodes which should not be used for patient or operator identification.
FDA-determined cause: Device Design
Action taken
Consignees of both devices were notified by Urgent Medical Device Correction letters sent via UPS Ground (receipt required) beginning 3/06/2013. The notice contained recommendation not to use EAN-13 barcode system and ensure barcodes used meet quality standards; use operator and patient lists (downloaded from a data manager) in order to check any decoded identification against the respective list. The notice directed forwarding of the notice if the product was further distributed. Customers were asked to complete the attached fax form. Contact the Customer Care Service Center, 24 hours a day, seven days a week at 1-800-440-3638 for Accu check questions and 1-800-428-4674 for CoaguChek questions about the information contained in this UMDC.
Details
- Recall number
- Z-1114-2013
- Initiated
- March 6, 2013
- Posted by FDA
- April 15, 2013
- Terminated
- January 10, 2018
- Product code
- GJS – Test, Time, Prothrombin
- Quantity
- 4467 both devices
- Distribution
- Nationwide Distribution.
- Recalling firm
- Roche Diagnostics Operations, Inc., Indianapolis, IN
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1113-2013 Class II | ACCU-CHEK Inform II Blood Glucose Monitoring System bar code scanner. Used with the…Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading… | 03/06/2013Terminated |