Recall Z-1115-2025

Beckman Coulter, Inc. · initiated January 3, 2024

Class II Completed

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument

Codes / lots: UDI/DI 15099590732103, serial numbers: 300208, 300201, 300184, 300185, 300169, 300225, 300175, 300172, 300221, 300209, 300211, 300212, 300213, 300214, 300215, 300210, 300174, 300226, 300204, 300216, 300217, 300170, 300220, 300176, 300188, 300199, 300200, 300177, 300222, 300224, 300173, 300181, 300202, 300191, 300223, 300171, 300186, 300190, 300203, 300178, 300195, 300189, 300198, 300227, 300228, 300168, 300194, 300205, 300206, 300207, 300197, 300218, 300179, 300182, 300192, 300219, 300193, 300183, 300180, 300187, 300196.

Reason for recall

A software error caused an increase in tray gripper motion errors that causes intermittent hard stops and a delay in patient results.

FDA-determined cause: Software design

Action taken

A field service engineer was dispatched to the customer site to perform the software update and to replace the PCB TSI-RSI board from 01/03/2024 through 12/26/2024.

Details

Recall number
Z-1115-2025
Initiated
January 3, 2024
Posted by FDA
February 6, 2025
Product code
JJE – Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Quantity
61 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Brazil, Croatia, France, Germany, Hungary, India, Ireland, Italy, New Zealand, Portugal, Slovakia, Spain, Switzerland, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland.
Recalling firm
Beckman Coulter, Inc., Chaska, MN

Official FDA recall record