Recall Z-1116-2015

Terumo Cardiovascular Systems Corporation · initiated February 6, 2015

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery.

Codes / lots: HX2 Temperature Management System, 120V Catalog # 809810 serial numbers: 00016, 0001 1116 HX2 Temperature Management System, 100V Catalog # 812231 serial numbers: 0001, 1002, PT0003, PT0004 HX2 Temperature Management System, 240V Catalog # 812240 serial numbers: 0001, 0002 HX2 Temperature Management System, Trial Kit Catalog # 809810TK serial numbers: 1001 1042

Reason for recall

Malfunctioning mixing valve and loss of cooling and/or heating capability. System goes to standby mode and shows error messages EOd (invalid mix valve measured range of motion) or EOE (valve unable to move or complete a motion). Mixing valves need to be replaced at regular intervals.

FDA-determined cause: Device Design

Action taken

On 2/6/2015, URGENT MEDICAL DEVICE RECALL notifications were mailed to the affected hospitals. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call customer service at 1-800-521-2818.

Details

Recall number
Z-1116-2015
Initiated
February 6, 2015
Posted by FDA
February 13, 2015
Terminated
October 17, 2016
Product code
DWC – Controller, Temperature, Cardiopulmonary Bypass
Quantity
103 units
Distribution
US (nationwide) including the states of AR, CA, CO, CT, FL, IA, IL, KS, KY, MD, MI, MO, MS, NY, OH, PA, TX, UT, WA, and WI and the country of Japan.
Recalling firm
Terumo Cardiovascular Systems Corporation, Ann Arbor, MI

Official FDA recall record