Recall Z-1125-2014

Synthes, Inc. · initiated February 5, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Synthes 5.0mm Variable Locking Screw This device is part of the Variable Angle Locking Compression Plate family and this technology included variable angle Combi holes in the plate shaft.The variable angle Combi hole provides the flexibility of axial compression and variable locking capacity throughout the length of the plate shaft.

Codes / lots: part no. 02.231.240, lot no. 823111

Reason for recall

Labeling correction: The package insert in one 5.0mm Variable Locking Screw package was a GP0006 (Anterior Cervical Vertebrae Plate System) insert, which should have been a GP2848 (4.5mm VA-LCP Curved Condylar Plate System insert).

FDA-determined cause: Employee error

Action taken

Synthes sent an Urgent Field Safety Notification/Medical Device Labeling Correction letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: Exchage the previously shipped package insert (GP0006) with the correct package insert (GP2848 provided with the notification). Review the package insert. Forward the Field Safety Notification to anyone in their facility that needs to be informed. If the package insert has been forwarded to another facility, contact that facility. Maintain awareness of this notice until all package inserts have been exchanged. Maintain a copy of the notice. Customers with questions were instructed to maintain a copy of the notice. For questions regarding this recall call 610-719-5000.

Details

Recall number
Z-1125-2014
Initiated
February 5, 2014
Posted by FDA
March 4, 2014
Terminated
August 31, 2015
Product code
JDP – Condylar Plate Fixation Implant
Quantity
45
Distribution
Nationwide Distribution including GA, CA, TX, MO, OH, PA, WA, AL, IN, AZ, MI, IA, TN, KY, NC, VA, NY, LA, IL, UT, and KS.
Recalling firm
Synthes, Inc., West Chester, PA