Recall Z-1130-2013

Lifescan Inc · initiated February 25, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

LifeScan brand OneTouch Select Control Solution, Part Number/Model #: 02168902. Testing of Blood Glucose Levels.

Codes / lots: 2AA2101, Exp. Date: 9-30-2014; 2AA2102, Exp. Date: 10-31-2-14

Reason for recall

When OneTouch Ultra and OneTouch Select test strips are tested with Ultra and Select control solution the meter may display a result outside the control solution test range printed on the strip vial label even though the system is working properly.

FDA-determined cause: Nonconforming Material/Component

Action taken

LifeScan sent an "URGENT MEDICAL DEVICE RECALL" letter dated February 25, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers may contact OneTouch Healthcare Professional Line at 866-461-0089 for questions about this notice.

Details

Recall number
Z-1130-2013
Initiated
February 25, 2013
Posted by FDA
April 17, 2013
Terminated
October 23, 2013
Product code
JJX – Single (Specified) Analyte Controls (Assayed And Unassayed)
Quantity
132,301 units total
Distribution
Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, 10, IL, IN, KS, KY, LA, MA, MD, ME, Ml, MN, MO, MS, MT, NC, NO, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SO, TN, TX, UT, VA, VI, VT, WA, WI,WV, and WY, and the countries of Russia and Ukraine.
Recalling firm
Lifescan Inc, Milpitas, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1129-2013 Class IILifeScan brand OneTouch Ultra Control Solution, Part Numbers/Model Numbers: 01045808…When OneTouch Ultra and OneTouch Select test strips are tested with Ultra and Select control solution the… 02/25/2013Terminated