Recall Z-1139-2007

Encore Medical, Lp · initiated August 15, 2006

Terminated

Product

Femoral Speed Block Size 10, Part/Catalog #800-01-369; positioning device included as part of the M.I.K.A. (Minimally Invasive Knee Arthroplasty) surgical kit; distributed by Encore Medical, L.P., Austin, TX 78758.

Codes / lots: M.I.K.A. Surgical Kits: MIKA-05-001, MIKA-05-002, MIKA-05-003, MIKA-05-004, MIKA-05-005, MIKA-05-006, MIKA-05-007, MIKA-05-008, MIKA-05-009, MIKA-05-010, MIKA-05-016, MIKA-05-018, MIKA-05-019, MIKA-05-020, MIKA-05-021, MIKA-05-022, MIKA-05-023, MIKA-05-024, MIKA-05-025, MIKA-05-026, MIKA-05-027, MIKA-05-028, MIKA-05-029, and MIKA-05-030.

Reason for recall

Complaints received that device used to prepare femur to receive femoral implant would split during surgery.

FDA-determined cause: Other

Action taken

Firm sent notification letter to consignees on 08/25/2006, along with replacement speedblocks. Consignees asked to trade out old speedblocks with the replacements.

Details

Recall number
Z-1139-2007
Initiated
August 15, 2006
Posted by FDA
August 7, 2007
Terminated
March 27, 2008
Product code
LKH – Antisera, Immunoperoxidase, Chlamydia Spp.
Quantity
24 units.
Distribution
Product distributed to direct accounts in USA and one consignee in Japan.
Recalling firm
Encore Medical, Lp, Austin, TX

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1135-2007 Femoral Speed Block Size 2, Part/Catalog #800-01-365; positioning device included as…Complaints received that device used to prepare femur to receive femoral implant would split during surgery. 08/15/2006Terminated
Z-1136-2007 Femoral Speed Block Size 4 Part/Catalog #800-01-366; positioning device included as part…Complaints received that device used to prepare femur to receive femoral implant would split during surgery. 08/15/2006Terminated
Z-1137-2007 Femoral Speed Block Size 6, Part/Catalog #800-01-367; positioning device included as…Complaints received that device used to prepare femur to receive femoral implant would split during surgery. 08/15/2006Terminated
Z-1138-2007 Femoral Speed Block Size 8, Part/Catalog #800-01-368; positioning device included as…Complaints received that device used to prepare femur to receive femoral implant would split during surgery. 08/15/2006Terminated
Z-1140-2007 Femoral Speed Block Size 12, Part/Catalog #800-01-370; positioning device included as…Complaints received that device used to prepare femur to receive femoral implant would split during surgery. 08/15/2006Terminated