Recall Z-1141-2025

HF Acquisition Co LLC · initiated December 11, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Biological, Bio52 HealthFirst Mail In Monitoring System, 52 Tests, Item: 1003600. Biological, Bio12 HealthFirst Mail In Monitoring System, 12 Tests, Item: 1003580

Codes / lots: Item/UDI-DI/Lot(Expiration): 1003600/10850065007820/ 2301060(11/30/2024) 2302261(11/30/2024) 2307261(12/31/2024 2312260(1/31/2025) 2313062(1/31/2025) 2331880(6/30/2025) 2401664(8/31/2025) 2414060(2/28/2026) 2414862(3/31/2026) 2414870(3/31/2026) 2416470(3/31/2026) 2416481(4/30/2026) 2416482(4/30/2026) 1003580/10850065007813/2421181(5/31/2026)

Reason for recall

Spore test strip monitoring system, designed for routine monitoring of sterilization processes, may have a printing error on the return envelope that may cause spore test results to be reported to the incorrect customer's account.

FDA-determined cause: Error in labeling

Action taken

On 12/11/2024, correction notices were mailed to customers asking them to do the following: 1. Check the box number on your Biological Weekly Monitoring kit for the affected lot numbers. The box number is located under the bar code on the outside of the box (see image below). 3. Check the serial number on all test strip return envelopes located in the box. This number is the last set of digits in the alphanumeric string located to the left of the bar code on the envelope. 4. Verify that the box number matches the last three digits of the serial number on the return envelope. 5. If the numbers do not match, you are hereby instructed to isolate your inventory of the above-listed lot numbers and dispose of it in regular trash. Firm will work with you on replacement product. 6. Ensure this letter is promptly brought to the attention of those responsible for managing the medical device inventory. Distribute this notification to all device users within your facility and network to ensure they are informed of this field correction letter. Please share this notification with any organizations that have received the potentially affected devices. 7. Complete and return the response form via email to quality@healthfirst.com Should you have any questions about this urgent field correction, please contact Customer Service at 855-476-1342 or email us at customerservice@healthfirst.com.

Details

Recall number
Z-1141-2025
Initiated
December 11, 2024
Posted by FDA
February 11, 2025
Product code
FRC – Indicator, Biological Sterilization Process
Quantity
4,021
Distribution
US: WI, FL, NY, CO, NJ, OH, IN, CA, WA, TX, AZ, MD, MN, NC, MI, MS, ID, TN, KS, SC, RI, LA, GA, PA, NH, UT, VA, MA, NV, MO, IL, SD, AR, OR, HI, NE, DC, CT, NM, ND, MT, OK, AL, VT, WY, KY, ME, IA, WV, AK
Recalling firm
HF Acquisition Co LLC, Mukilteo, WA

Official FDA recall record