Recall Z-1146-2013

Edwards Lifesciences, LLC · initiated January 30, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Pressure Monitoring sets with VAMP Plus closed blood sampling system model # T001691A, PXVP0566, PXVP2260, PXVP0826, PXVP0827, PXVP2284AT3, PXVP23X3AT3, T398807D, T100605B, T001762A, T391T00A, VP2. Blood sampling system used on patients requiring periodic withdrawal of blood samples.

Codes / lots: 59184554, 59187809, 59191725, 59221801, 59227844, 59238613, 59326333, 59356987, 59401041, 59222349, 59227178, 59234808, 59284693, 59328047, 59270408, 59284617, 59276238, 59401734, 59270380, 59319038, 59306624, 59390840, 59195810, 59331224, 59397181, 59291446, 59272771.

Reason for recall

The device is being recalled because the firm has received multiple complaints related to incorrect assembly that may result in dilution of blood samples.

FDA-determined cause: Process change control

Action taken

The firm initiated a recall in Europe via customer letters starting January 30, 2013 to notify the recall of model number T001691A lot number 59326333. The recall was expanded and a worldwide recall via customer letters was initiated on February 11, 2013. The letter identified the product, problem, and actions to be taken by the customers.

Details

Recall number
Z-1146-2013
Initiated
January 30, 2013
Posted by FDA
April 18, 2013
Terminated
March 10, 2016
Product code
KRA – Catheter, Continuous Flush
Quantity
30,467
Distribution
Worldwide Distribution-USA (nationwide) including the states of AR, CA, FL, HI, ID, IN, KS, KY, LA, MA, MI, MN, MO, NC, ND, NM, NV, NY, OH, OK, PA, SD, TN, TX, VA, WA, and WI) and Europe.
Recalling firm
Edwards Lifesciences, LLC, Irvine, CA

Official FDA recall record