Recall Z-1146-2014
Horiba Instruments, Inc dba Horiba Medical · initiated January 20, 2014
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
ABX PENTRA Multical. ABX PENTRA N Control. ABX PENTRA P Control. The ABX PENTRA 400 and PC200 are benchtop clinical chemistry analyzers using two measuring principals: absorbance and ion selective electrodes. The ABX PENTRA Multical is a lyophilized human serum calibrator with chemical additives and materials of biological origin. The assigned values of the calibrator components are given in the enclosed annex, ensuring optimal calibration of the appropriate HORIBA ABX SAS methods on the ABX PENTRA Clinical Chemistry Analyzer. This calibrator is provided in ten vials of 3 ml.
Codes / lots: Model Numbers:Multical - A11A01652, N Control - A11A01653, P Control - A11A01654. Lot Numbers: MultiCal: 1204101 and 1304901, N Control: 1300801, P Control: 1300601. Expiration Date (s): MultiCal: 1204101 Expires on 05/31/14, MultiCal: 1304901 Expires on 07/15/15, N Control: 1300801 Expires on 11/05/14, P Control: 1300601 Expires on 11/05/14.
Reason for recall
HORIBA Medical is informing all ABX PENTRA MultiCal, N Control, and P Control customers running the Creatinine Rate Blank Method assay on the ABX PENTRA 400 and PC200 that the Target Values for Creatinine Rate Blank Method have been updated. This issue is only applicable to Creatinine Rate Blank Method values reported on the ABX PENTRA 400 and PC200 Chemistry Analyzers only. No other analyzers
FDA-determined cause: Under Investigation by firm
Action taken
HORIBA Medical initiated this recall by sending customers recall communication via FedEx week of January 20, 2014. The recall communication was a Field Safety Notice letter and Acknowledgement Form. The letter, dated 01/20/2014, titled "URGENT SAFETY NOTICE", informed customers of the recall by providing product description and codes, issue, impact, actiona/resolution, and contact information. The letter contained the following attachments: Field Safety Notice Acknowledgement Form and Updated Annexes with re-assigned values.
Details
- Recall number
- Z-1146-2014
- Initiated
- January 20, 2014
- Posted by FDA
- March 5, 2014
- Terminated
- December 5, 2014
- Product code
- JIX – Calibrator, Multi-Analyte Mixture
- Quantity
- 2,023 units
- Distribution
- United States nationwide distribution.
- Recalling firm
- Horiba Instruments, Inc dba Horiba Medical, Irvine, CA