Recall Z-1147-2020

Stryker Corporation · initiated July 19, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Stryker, Omega 4.75 mm,. PEEK Knotless Anchor System, Double-Double, REF 3910-500-472, UDI 07613327464832, Sterile R, RX Only CE 0197

Codes / lots: 19113AG2, 19130AG2, 19134AG2, 19115AG2, 19131AG2, 19133AG2, 19154AG2,19155AG2, 19156AG2, 19157AG2, 19149AG2.

Reason for recall

Complaints have been received regarding difficulty in removing the driver from the screw. High driver removal forces can lead to the screw pulling out still attached to the driver. This could result in an additional anchor insertion site in the bone being used.

FDA-determined cause: Nonconforming Material/Component

Action taken

On July 23, 2019, a "Urgent Medical Device Recall Notification" was emailed to sales representatives. In addition to informing consignees on the cause of the recall, the following action was requested: 1. Check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your accounts. Response is required by August 15, 2019. 2.If affected product is found in your inventory, segregate the product and call Stryker customer service at 1-866-596-2022 or email smwest@stryker.com to arrange for product return and issuance of credit or replacement (upon availability). Also fill out Attachment A and return to endorecall@stryker.com. 3.If affected product is found at account(s), please complete both steps below: a.Provide attached memo (Attachment B) to the account(s) to acknowledge the removal of devices and send completed memo back to endorecall@stryker.com. b.Segregate the product and call Stryker customer service at 1-866-596-2022 or email smwest@stryker.com to arrange for product return and issuance of credit or replacement (upon availability). 4.If no product is found in your inventory or at your account(s), complete acknowledgement form on Attachment A and return it back to endorecall@stryker.com.

Details

Recall number
Z-1147-2020
Initiated
July 19, 2019
Posted by FDA
March 2, 2020
Terminated
February 2, 2024
Product code
MBI – Fastener, Fixation, Nondegradable, Soft Tissue
Quantity
73 devices
Distribution
US: GA, NJ, NY, KY, MD, TX, UT, AZ, IL, ID, CO, OH, FL, AK, CA, NV, AR, ME, WI
Recalling firm
Stryker Corporation, San Jose, CA