Recall Z-1149-2023

Siemens Healthcare Diagnostics, Inc. · initiated January 9, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Atellica IM Folate 140 test kit- For in vitro diagnostic use in the quantitative determination of folate in human serum and red blood cells. Siemens Material Number (SMN): 10995572

Codes / lots: (01)00630414598963(10)15308337(17)20230320 (01)00630414598963(10)27949339(17)20230529 (01)00630414598963(10)41143343(17)20230630 (01)00630414598963(10)41144343(17)20230630 (01)00630414598963(10)62969345(17)20230726 Lot Numbers: 15308337, 27949339, 41143343, 41144343, 62969345 and higher. Affects all current and future lots of the Atellica IM Folate assay until the Instructions for Use are updated.

Reason for recall

Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.

FDA-determined cause: Under Investigation by firm

Action taken

Siemens Healthineers issued Urgent Medical Device Corrections (UMDC AIMC 23-03.A-1.US and AIMC 23-03.A-2.US) to US customers and Urgent Field Safety Notices (UFSN AIMC 23-03.A-1.OUS and AIMC 23-03.A-2.OUS) to customers outside the United States. The UMDCs and UFSNs advise customers to use the appropriate FolSerum calibration instructions when testing serum samples with Atellica IM Folate kit lots ending in 337 and above or ADVIA Centaur Folate kit lots ending in 336 and above. Letter states reason for recall, health risk and action to take: Follow the instructions in this Urgent Medical Device Correction until the Atellica IM Folate/ADVIA Centaur Folate Instructions for Use are updated. o Before updating to the new reagent lot, ensure you have processed all serum samples necessary for lot-to-lot comparisons using your existing inventory. o Once you have updated to the new reagent lot, you will not be able to process serum samples with the previous reagent lot. o If your laboratory runs both serum and whole blood sample types, then both assays must be calibrated with the new reagent lot. o If your laboratory runs one sample type, only calibrate the sample type you use If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.

Details

Recall number
Z-1149-2023
Initiated
January 9, 2023
Posted by FDA
February 21, 2023
Product code
CGN – Acid, Folic, Radioimmunoassay
Quantity
13,843 kits
Distribution
Nationwide Foreign: Albania¿¿¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿ Belgium¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Bulgaria¿¿¿¿¿¿¿ Croatia¿¿¿¿¿¿¿¿ Cura¿ao,¿St¿Eus Cyprus¿¿¿¿¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ Estonia¿¿¿¿¿¿¿¿ Finland¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Georgia¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Hungary¿¿¿¿¿¿¿¿ Iran¿¿¿¿¿¿¿¿¿¿¿ Ireland¿¿¿¿¿¿¿¿ Israel¿¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Jordan¿¿¿¿¿¿¿¿¿…
Recalling firm
Siemens Healthcare Diagnostics, Inc., East Walpole, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1150-2023 Class IIAtellica IM Folate 700 test kit- For in vitro diagnostic use in the quantitative…Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used… 01/09/2023Open, Classified
Z-1151-2023 Class IIAtellica IM Folate 700 test kit (REF)-For in vitro diagnostic use in the quantitative…Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used… 01/09/2023Open, Classified
Z-1152-2023 Class IIADVIA Centaur Folate 100 test kit-For in vitro diagnostic use in the quantitative…Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used… 01/09/2023Open, Classified
Z-1153-2023 Class IIADVIA Centaur Folate 500 test kit- For in vitro diagnostic use in the quantitative…Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used… 01/09/2023Open, Classified
Z-1154-2023 Class IIADVIA Centaur Folate 500 test kit (REF)-For in vitro diagnostic use in the quantitative…Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used… 01/09/2023Open, Classified