Recall Z-1156-2007

Datascope Corporation · initiated May 7, 2007

Terminated

Product

Precision Bipolar Device ClearGlide EVH Small (endoscopic vessel harvesting system) The kit included three Precision BiPolar Devices, an Optical Vessel Dissector, a Small Ultra Retractor and a Vessel Dissector. Datascope Cardiac Assist Sterile R Manufactured for: Datascope Cardiac Assist 15 Law Drive Fairfield, NJ 07004

Codes / lots: Part number KTV 15 Lot numbers: 12499, Exp. 4/21/09; 12500, Exp. 4/22/09; 12501, Exp. 4/27/09; 12504, Exp. 5/04/09 and 12505, Exp. 5/13/09.

Reason for recall

Datascope has experienced a higher than usual number of reports involving the Precision Bipolar Device. Reports include the locking or sticking of the buttons of the device which operate the jaws, particularly when a side load is placed on the shaft of the device.

FDA-determined cause: Other

Action taken

Letters were sent to all hospitals, (contact: Hospital Administrators) the week of May 7, 2007 (certified mail, Return Receipt requested) with Urgent Medical Device Field Correction. The letter requests that all potential users of the devices at the institution be mdae aware of the situation. The letter instructs clinicians what to do if they experience problems with locking buttons of the Precision Device and also if the device is clamped closed on a vein branch and cannot be opened. Institutions are also requested to sign and return a postal card acknowledging receipt of the Field Correct notification and Datascope provides telephone numbers for customer service.

Details

Recall number
Z-1156-2007
Initiated
May 7, 2007
Posted by FDA
August 9, 2007
Terminated
March 4, 2008
Product code
GEI – Electrosurgical, Cutting & Coagulation & Accessories
Quantity
1008
Distribution
Worldwide, including USA, Belgium, the Netherlands, and Germany.
Recalling firm
Datascope Corporation, Fairfield, NJ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1157-2007 The ClearGlide EVH small with Scissors. The kit includes three Precision Bipolar…Datascope has experienced a higher than usual number of reports involving the Precision Bipolar Device.… 05/07/2007Terminated
Z-1158-2007 Precision Bipolar Device 6 pack Consists of a six-pack of the Precision Bipolar Device.…Datascope has experienced a higher than usual number of reports involving the Precision Bipolar Device.… 05/07/2007Terminated