Recall Z-1159-2008
Bio-Rad Laboratories · initiated November 13, 2007
Product
BIOPLEX 2200 MULTI-ANALYTE DETECTION SYSTEM: a) 100 test - Catalog No. 665-1150; and b) the 5,000 tests - Catalog No. 665-1155 - manufactured by Bio-Rad Laboratories in Redmond, WA. The product is used as an aid in the diagnosis of systemic autoimmune diseases.
Codes / lots: Eleven (11) Lots, as follows: Lot Numbers: 931396 Exp. 9/30/06; 935424 Exp. 3/31/07; 941886 Exp. 3/31/07; 935682 Exp. 4/30/07; 937351 Exp. 6/30/07; 941343 Exp. 3/31/08; 941428 Exp. 4/30/08; 945246 Exp. 9/30/08; 945875 Exp. 12/31/08; 945248 Exp. 10/31/08 and 945875 Exp. 12/31/08 ****
Reason for recall
False negative results due to reagent packs exhibiting low signal.
FDA-determined cause: Nonconforming Material/Component
Action taken
On 11/13/2007, the firm faxed Voluntary Field Correction Letters (which are dated 11/9/2007) to its consignees and also contacted them via telephone. The Voluntary Field Correction Letters: a) informed them of the potential for false negative results; b) recommended that they perform QC testing daily with each reagent pack; c) provided a new insert for the Instructions for Use; and d) provided a Response Sheet for each consignee to complete and return to the firm.
Details
- Recall number
- Z-1159-2008
- Initiated
- November 13, 2007
- Posted by FDA
- June 11, 2008
- Terminated
- May 23, 2011
- Product code
- LKJ – Antinuclear Antibody, Antigen, Control
- Quantity
- 8,804 kits
- Distribution
- Worldwide Distribution -- USA, Canada, France, United Kingdom, Italy, Germany, South Africa, and Switzerland.
- Recalling firm
- Bio-Rad Laboratories, Redmond, WA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1156-2008 | BIOPLEX 2200 SYPHILIS IgG Kit: 100 tests - Catalog No. 665-1450 - manufactured by…False negative results due to reagent packs exhibiting low signal. | 11/13/2007Terminated |
| Z-1157-2008 | BIOPLEX 2200 EBV IgG KIT, CALIBRATORS, AND CONTROLS: 100 tests - Catalog No. 665-1250 -…False negative results due to reagent packs exhibiting low signal. | 11/13/2007Terminated |
| Z-1158-2008 | BIOPLEX 2200 EBV IgM KIT, CALIBRATORS, AND CONTROLS: 100 tests - Catalog No. 665-1350 -…False negative results due to reagent packs exhibiting low signal. | 11/13/2007Terminated |