Recall Z-1163-2025

Microbiologics Inc · initiated January 20, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Microbiologics Multi-Strain Human Papillomavirus (HPV) Control Panel (Inactivated Pellet), REF 8234; in-vitro diagnostic test

Codes / lots: UDI/DI 70845357043077, Lot Number 8234-22

Reason for recall

Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 L1) was low enough to cause QC failures, no detection, or invalid results on some instruments and assays.

FDA-determined cause: Under Investigation by firm

Action taken

Microbiologics issued and Urgent Medical Device Recall notice to its consignees on 01/20/2025 via email. The notice explained the issue and requested the following actions be taken: INSTRUCTIONS: 1. REVIEW lab procedures to understand how this information affects your usage 2. DISCARD affected product. 3. COMPLETE the response form provided. 4. RETURN the response form to recall@microbiologics.com 5. KEEP this letter for your records. 6. NOTIFY other facilities or customers to whom the product may have been distributed or transferred. 7. CONTACT Microbiologics if a replacement kit is needed. Please contact our Recall Support team at 320.229.7080 or recall@microbiologics.com with questions or concerns.

Details

Recall number
Z-1163-2025
Initiated
January 20, 2025
Posted by FDA
February 12, 2025
Product code
OHQ – Multi-Analyte Controls Unassayed
Quantity
16 units
Distribution
Worldwide distribution - US Nationwide and the countries of Netherlands, Sweden, UK, France, Australia, New Zealand.
Recalling firm
Microbiologics Inc, Saint Cloud, MN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1164-2025 Class IIMicrobiologics Multi-Strain Human Papillomavirus (HPV) Verification Panel (Inactivated…Internal quality control procedures were not followed correctly for the testing/release of three lots of… 01/20/2025Open, Classified