Recall Z-1169-2013
Hospira Inc. · initiated March 18, 2013
Class I – dangerous or defective products that could cause serious health problems or death
Product
The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar Pump can be powered by AC mains adaptor, rechargeable battery pack, docking station, or two disposable AA alkaline batteries. When powered by batteries, The GemStar Pump is ideal for ambulatory patients. The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required.
Codes / lots: List Numbers: 13000, 13100, 13150
Reason for recall
Pump shutting off during use without warning.
FDA-determined cause: Device Design
Action taken
Hospira sent an Important Device Information letter dated March 18, 2013, to all affected customer. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inspect the affected product for signs of leakage, corrosion or other damage prior to each use. In addition, each time the batteries are replaced the battery compartment should be inspected for damage. If a device exhibits damage caused by leakage, it should immediately be removed from clinical service and Hospira should be contacted at 1-800-441-4100 to arrange for return of the device for repair. If the product has been further distributed they should notify their consignees and confirm they have done so by returning the reply form to Stericycle. For questions regarding this recall call 224-212-4892.
Details
- Recall number
- Z-1169-2013
- Initiated
- March 18, 2013
- Posted by FDA
- April 30, 2013
- Terminated
- March 10, 2017
- Product code
- FRN – Pump, Infusion
- Quantity
- List Number 13000: 25,119 pumps; List Number 13100: 185 pumps; List Number 13150: 13,497 pumps
- Distribution
- Worldwide Distribution - USA (nationwide) and internationally to Australia, New Zealand, China, Hong Kong, Taiwan, Japan, Korea, Philippines, Thailand, Malaysia, Canada, Austria, Bahrain, Belgium, Croatia, Denmark, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Iceland, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Netherlands, Norway, Oman, Portugal…
- Recalling firm
- Hospira Inc., Lake Forest, IL
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1170-2013 Class I | The GemStar Pump is a small and lightweight, single channel infusion device designed for…Pump shutting off during use without warning. | 03/18/2013Terminated |