Recall Z-1171-2016

Sekisui Diagnostics P.E.I. Inc. · initiated January 22, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Sekure Chemistry Sekisui Diagnostics Acetaminophen L3K Assay, In-vitro Diagnostic Catalog Number: 506-10: R1: 1 x 10ml, R2: 2 x 10ml Catalog Number: 506-30: R1: 3 x 10ml, R2: 6 x 10ml In-Vitro quantitative measurement of Acetaminophen in serum and plasma used m the diagnosis and treatment of acetaminophen overdose toxicity.

Codes / lots: Catalog Number: 506-10 Lot Codes 47677 47694 47649 47650 47662 47663 48253 48385 48386 Catalog Number: 506-30 Lot Codes: 47695 48294

Reason for recall

Crystal formation in the Color Reagent (R2) when left on board automated analyzers causing samples to under recover

FDA-determined cause: Device Design

Action taken

Sekisui sent an Important Field Correction - Customer Notification letter dated January 22, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1. Discard all reagent in house with a current expiration date of 2016-09. 2. Update the expiration on all remaining kits on hand with the new expiration dates provided 3. When adding additional reagent to the analyzer, a new reagent wedge should be used. 4. If crystals are observed within reagent stored on-board, reagent should be discarded and replaced with fresh reagent. 5. Complete, sign and return the attached CONFIRMATION OF NOTIFICATION form by FAX to the number provided. This will indicate receipt of this Field Correction, act as confirm action of compliance with the actions,and allow for issuance of replacement material for expired product on hand (lots 47677, 47694, 47695). Questions contact Technical Services at 1-800-565-0265 or via email at PEIDiagnosticTechnical@sekisui-dx.com

Details

Recall number
Z-1171-2016
Initiated
January 22, 2016
Terminated
August 9, 2016
Product code
LDP – Colorimetry, Acetaminophen
Quantity
14,407 kits.
Distribution
Worldwide Distribution - US Distribution to the states of : AZ, CA, FL, IA, ID, IL, MA, , MN, MT and ND., and to the countries of Canada, France and Hong Kong.
Recalling firm
Sekisui Diagnostics P.E.I. Inc., Charlottetown