Recall Z-1172-2025

Augustine Temperature Management, LLC · initiated February 15, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A138 used in conjunction with Warming Blankets and Mattresses.

Codes / lots: UDI-DI 00850011479674, Lot Number 2339

Reason for recall

Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .

FDA-determined cause: Device Design

Action taken

Augustine Temperature Management contacted their consignees by visit beginning on 02/15/2024 to replace the affected cables. The firm collected email confirmation that the original cables had either been returned or destroyed. If you have any questions, contact the Product Manager at 952-465-3526 or email:

Details

Recall number
Z-1172-2025
Initiated
February 15, 2024
Posted by FDA
February 14, 2025
Product code
DWJ – System, Thermal Regulating
Quantity
7 unit
Distribution
US distribution to states of: AL, TX, NV
Recalling firm
Augustine Temperature Management, LLC, Minnetonka, MN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1171-2025 Class IIAugustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A136…Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact . 02/15/2024Open, Classified