Recall Z-1173-2020
Arrow International Inc · initiated January 15, 2020
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
EPIDURAL CATHETERIZATION KIT, Product Code ALZANO-05400-B
Codes / lots: Lots: 71F19B0681 71F19C0035 71F19D2827 71F19F2028 71F19H1633 71F19K1574
Reason for recall
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
FDA-determined cause: Other
Action taken
An Urgent Medical Device Recall notification letter dated 1/15/20 was sent to customers. Our records indicate you have received products that are subject to this field action. We are now notifying our customers to take the following actions: 1. If you have affected stock, immediately identify product, discontinue use and quarantine any products with the product codes and lot numbers listed above. 2. To acknowledge receipt of this Urgent Medical Device Recall Notice or to return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document your receipt of this letter.
Details
- Recall number
- Z-1173-2020
- Initiated
- January 15, 2020
- Posted by FDA
- January 24, 2020
- Terminated
- March 17, 2023
- Product code
- CAZ – Anesthesia Conduction Kit
- Quantity
- 343,105 total
- Distribution
- The products were distributed to the following US states: AK, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY.
- Recalling firm
- Arrow International Inc, Reading, PA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1170-2020 Class II | FlexTip Plus Combined Spinal Epidural Catheterization Set, Product Code AA-05400-BThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1171-2020 Class II | FlexTip Plus Epidural Catheterization Set, Product Code AA-05400-EThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1172-2020 Class II | FlexTip Plus Epidural Catheterization Kit, Product Code AK-05501There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1174-2020 Class II | FlexTip Plus Epidural Catheterization Kit, Product Code ASK-02220-SRHThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1175-2020 Class II | TheraCath Epidural Catheterization Kit, Product Code ASK-05001-SLR1There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1176-2020 Class II | FlexTip Plus Epidural Catheterization Kit, Product Code ASK-05500-TMThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1177-2020 Class II | FlexTip Plus Epidural Catheterization Kit, Product Code AK-05500There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1178-2020 Class II | FlexTip Plus Combined Spinal Epidural Catheterization Kit, Product Code ASK-05500-BIDThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1179-2020 Class II | FlexTip Plus Epidural Catheterization Kit, Product Code ASK-05400-CA1There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1180-2020 Class II | FlexTip Plus Epidural Catheterization Kit, Product Code ASK-05501-SHThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1181-2020 Class II | Epidural Catheterization Kit with FlexTip Plus, Open Tip, Single- Port Catheter, Product…There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1182-2020 Class II | FlexTip Plus Epidural Catheterization Kit, Product Code ASK-05503-BIDThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1183-2020 Class II | FlexTip Plus Combined Spinal Epidural Catheterization Kit, Product Code ASK-05560-TG1There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1184-2020 Class II | FlexTip Plus Combined Spinal Epidural Catheterization Kit, Product Code ASK-05560-WHThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1185-2020 Class II | FlexTip Plus Epidural Catheterization Kit, Product Code ASK-17019-MSCThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1186-2020 Class II | EPIDURAL CATHETERIZATION KIT, Product Code AT-05501-LEOThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1187-2020 Class II | EPIDURAL CATHETERIZATION KIT, Product Code AT-05501-LINThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1188-2020 Class II | EPIDURAL CATHETERIZATION KIT, Product Code BE-05400B-ETTELThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1189-2020 Class II | EPIDURAL CATHETERIZATION KIT, Product Code BE-05400-DCHHThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1190-2020 Class II | EPIDURAL CATHETERIZATION KIT, Product Code BE-05400-DCSHOThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1191-2020 Class II | EPIDURAL CATHETERIZATION KIT, Product Code BE-080180-BXLThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1192-2020 Class II | EPIDURAL CATHETERIZATION KIT, Product Code BE-080180-CHHThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1193-2020 Class II | FlexTip Plus Epidural Catheterization Kit, Product Code CA-02220There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1194-2020 Class II | EPIDURAL CATHETERIZATION KIT, Product Code CZ-05400-EPIThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1195-2020 Class II | EPIDURAL CATHETERIZATION KIT, Product Code DE-05400D-BOThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1196-2020 Class II | FlexTip Plus Epidural Catheterization Set for Pediatric Lumbar Placement, Product Code…There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1197-2020 Class II | EPIDURAL CATHETERIZATION KIT, Product Code FR-05501-04There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1198-2020 Class II | EPIDURAL CATHETERIZATION KIT, Product Code FR-05501-10There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1199-2020 Class II | EPIDURAL CATHETERIZATION KIT, Product Code FR-05501-12There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |
| Z-1200-2020 Class II | EPIDURAL CATHETERIZATION KIT, Product Code IT-05400-DCThere is the potential for various failures with the LOR (loss of resistance) syringes included in the kits… | 01/15/2020Terminated |