Recall Z-1175-2008

California Medical Laboratories Inc · initiated January 24, 2008

Terminated

Product

CalMed Pericardial Sump, with 1/8"/0.3175 cm Connector Single Use Only. Do Not Reuse. REF SU-20601; CalMed Laboratories, Costa Mesa, CA 92626

Codes / lots: Catalog Number SU-20601, Lot Number S03482

Reason for recall

Fragment of material: The product may have a loose, stainless steel stringer flash located at the distal tip of the product.

FDA-determined cause: Process control

Action taken

CalMed emailed letters to consignees on 1/24/2008 notifying customers of the recall. The letter advises that that the following Medical Devices are subject to an immediate Voluntary Recall for Correction: Pericardial Sump (Catalog #SU-20601), Pericardial Sump (Catalog #SU-20602), and Malleable Pericardial Sump (Catalog #SU-20802). It is asked that further distribution or use of any remaining product should cease immediately. Where appropriate, the firm asks that consignees notify their own customers who received the product of the same. The Product is being recalled due to a potentially loose, stainless steel stringer flash located at the distal tip of the Product. Customers are asked to contact the firm's Quality Assurance Manager immediately for purposes of issuing a Returned Goods Authorization and coordinating the return of the Subject Medical Devices such that they may be inspected and nonconforming product removed. A response is requested via e-mail.

Details

Recall number
Z-1175-2008
Initiated
January 24, 2008
Posted by FDA
June 11, 2008
Terminated
March 5, 2013
Product code
DTS – Sucker, Cardiotomy Return, Cardiopulmonary Bypass
Quantity
31,124 units of all models
Distribution
Worldwide Distribution - USA, Turkey, Brazil, Germany, Italy, Spain, Philippines, South Africa, Japan, Israel, Saudi Arabia, United Arab Emirates, India, Ireland, China, Hungary, Poland, Taiwan, Belgium, Pakistan, Lithuania, Greece, Canada, Syria, Israel, Austria, Denmark, Bangladesh, Costa Rica, Mexico, Spain, and Sultanate of Oman.
Recalling firm
California Medical Laboratories Inc, Costa Mesa, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1176-2008 CalMed Pericardial Sump, with 1/4"/0.635 cm Connector, Single Use Only. Do Not Reuse…Fragment of material: The product may have a loose, stainless steel stringer flash located at the distal tip… 01/24/2008Terminated
Z-1177-2008 CalMed Malleable Pericardial Sump, Single Use Only. Do Not Reuse. REF SU-20802; CalMed…Fragment of material: The product may have a loose, stainless steel stringer flash located at the distal tip… 01/24/2008Terminated