Recall Z-1184-2023
Stryker Corporation · initiated January 25, 2023
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Finished Goods Box of 50 Inserts and Scanning Labels for Medivac Guardian 2L Catalog Number: FG 12010. Intended adjunct in the estimation of blood loss.
Codes / lots: UDI-DI: 10859506006125 Lot Numbers: M2110004, M2201002, M2202001, M2203002, M2206001
Reason for recall
Triton Canister Software, Insert & Scanning Label may cause the Triton Canister System to output inaccurate blood loss estimates can lead to confusion and could potentially result in either delayed recognition or prolongation of postpartum hemorrhage (PPH), or require more aggressive treatment for PPH.
FDA-determined cause: Under Investigation by firm
Action taken
Stryker issued Urgent Medical Device Correction letter to Risk Manager, OR Director, Materials Manager dated Jan 25, 2023. Letter states reason for recall, health risk and action to take: Steps: 1. Immediately check your internal inventory to locate any affected Finished Goods Inserts and Scanning labels (FG 12010 and FG 12022). Quarantine and discontinue use of all affected products. 2. A Stryker representative will contact your facility to assist updating the Triton Canister software. 3. Return the enclosed Business Reply Form (BRF) even if you don t have any affected product on hand, via email to instruments.recalls@stryker.com. 4. If the BRF indicates that recalled product is currently on hand, a shipping label will be provided to return recalled product. Upon receipt of the recalled product, Stryker will arrange replacements. 5. Maintain awareness of this communication internally until all required actions have been completed within your facility. Contact: Sylvia Page | Stryker Instruments Recall Coordinator 269-501-8375 |Instruments.recalls@stryker.com
Details
- Recall number
- Z-1184-2023
- Initiated
- January 25, 2023
- Posted by FDA
- February 27, 2023
- Product code
- PBZ – Image Processing Device For Estimation Of External Blood Loss
- Quantity
- 101 units
- Distribution
- US Nationwide distribution.
- Recalling firm
- Stryker Corporation, Portage, MI
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1185-2023 Class II | Finished Goods Box of 50 Inserts and Scanning Labels for Medivac Guardian 3L Catalog…Triton Canister Software, Insert & Scanning Label may cause the Triton Canister System to output inaccurate… | 01/25/2023Open, Classified |
| Z-1186-2023 Class II | Triton Canister Software Catalog Number: GAUS-2. Intended adjunct in the estimation of…Triton Canister Software, Insert & Scanning Label may cause the Triton Canister System to output inaccurate… | 01/25/2023Open, Classified |