Recall Z-1186-2016

Sorin Group USA, Inc. · initiated August 15, 2012

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Stockert Heater-Cooler 3T; Item Number 16-02-8, H3T/240V/60Hz ; 16-02-82, H3T208V/60Hz ; 16-02-85, H3T/120V/60Hz. Device Listing E572259

Codes / lots: All Serial Numbers for Item Number 16-02-81, 16-02-82, 16-02-85;

Reason for recall

On August 15, 2012, Sorin informed consignees via letter of changes made to create Version 12 of the Heater-Cooler 3T IFU. The revised IFU included enhancements to achieve water quality that meets the standard for German drinking water.

FDA-determined cause: Device Design

Action taken

Sorin Germany, via Sorin Colorado, distributed an important information letter to its US customers on August 15, 2012 informing them about an update of the cleaning and infection procedure for the U.S. market that were included in Revision 12 of the IFU.

Details

Recall number
Z-1186-2016
Initiated
August 15, 2012
Posted by FDA
March 17, 2016
Terminated
May 26, 2017
Product code
DWC – Controller, Temperature, Cardiopulmonary Bypass
Quantity
1125 units
Distribution
Worldwide Distribution - US Nationwide, VA/govt/military
Recalling firm
Sorin Group USA, Inc., Arvada, CO