Recall Z-1187-2021

Tosoh Bioscience Inc · initiated February 23, 2012

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Thyroglobulin Antibody (TgAB) Calibrator - Product Usage: intended for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of anti-thyroglobulin antibodies (TgAb) in human serum on specific TOSOH AIA System analyzers. Measurements obtained by this device aid in the diagnosis of autoimmune thyroid disorders, such as Graves disease and Hashimotos disease. Comment to IU: Please note that the Calibrator set is not part of the IU but required for the assay.

Codes / lots: Part Number 020391; Lot # B834868, Lot # BX34869, Lot # C234871, Lot # C434872 & Lot # C534873.

Reason for recall

Some combinations of Calibrator lots and AlA PACKS resulted in Calibrator #6 not giving a value, making the calibration unacceptable.

FDA-determined cause: Nonconforming Material/Component

Action taken

On 7/24/2012 the firm sent recall notificaiton letters to their customers.

Details

Recall number
Z-1187-2021
Initiated
February 23, 2012
Terminated
July 8, 2021
Product code
JZO – System, Test, Thyroid Autoantibody
Quantity
34 boxess/2 sets/6 bottles
Distribution
US Nationwide distribution in the states of CA, FL, GA, NY, OH & OR.
Recalling firm
Tosoh Bioscience Inc, Grove City, OH

Official FDA recall record