Recall Z-1188-2021

Siemens Healthcare Diagnostics, Inc. · initiated February 3, 2021

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Dimension Vista HDLC - colorimetric method, lipoproteins - Product Usage: used as an aid in the diagnosis of lipid disorders (such as diabetes mellitus), various liver and renal diseases, and in the assessment of risk for atherosclerosis and cardiovascular disease.

Codes / lots: Product: DV HDLC; Lot #: 20062BA; UDI Number: 0084276802043820062BA21030210464340840

Reason for recall

Siemens Healthcare Diagnostics Inc. has confirmed the potential for erroneous results due to discolored reagent observed in some wells 5 and 6 of Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridges lot 20062BA.

FDA-determined cause: Under Investigation by firm

Action taken

Urgent Medical Device Correction (UMDC # VC-21-01.A.US) and Urgent Field Safety Notice (UFSN # VC-21-01.A.OUS) were issued on January 26, 2021 to all Dimension Vista customers who received the Dimension Vista HDLC flex reagent lot 20062BA to notify them of the issue, the potential risk to health, and actions to be taken by the customer.

Details

Recall number
Z-1188-2021
Initiated
February 3, 2021
Product code
JHM – Colorimetric Method, Lipoproteins
Quantity
4174 devices (Domestic: 2836 devices, Foreign: 1338 devices)
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY, and the countries of Austria, Belgium, Canada, Denmark, France, Germany, Italy, Netherlands, Portugal, SA, Slovakia, Slovenia, Spain…
Recalling firm
Siemens Healthcare Diagnostics, Inc., Newark, DE

Official FDA recall record