Recall Z-1190-2008

Steris Corp · initiated February 19, 2008

Terminated

Product

Steris C1160 Universal Flexible Processing Tray, for use with STERIS SYSTEM 1 Processor, used for sterilization of flexible endoscopes.

Codes / lots: All codes

Reason for recall

Failure to effectively sterilize: Under certain conditions, the design/operation of the C1160 Tray may compromise the ability of the SYSTEM 1 processor to correctly assess the state of the high pressure pump during the Sterile Cycle. The L6 sensor fails to alarm when the high pressure pump is not working properly.

FDA-determined cause: Device Design

Action taken

Steris sent an Urgent Safety Alert/Recall letter dated 2/8/2008, to all customers. The letter stated that any device sterilized in the C1160 Tray should not be used clinically prior to a completion of a successful diagnostic cycle. Until further notice, you must run a diagnostic cycle after each sterile cycle involving a C1160 tray or alternately you must discontinue use of these trays. Steris service technicians will schedule visits to correct the problem.

Details

Recall number
Z-1190-2008
Initiated
February 19, 2008
Posted by FDA
August 5, 2008
Terminated
February 24, 2012
Product code
MED – Sterilant, Medical Devices
Quantity
28,766 units
Distribution
Worldwide Distribution, including USA, Puerto Rico, Canada, and Japan.
Recalling firm
Steris Corp, Mentor, OH

Official FDA recall record